吲哚菁绿联合Firefly技术用于BABA机器人甲状腺切除术中甲状旁腺定位
该病例对照研究评估近红外光诱导的吲哚菁绿荧光在BABA机器人甲状腺切除术中识别并保护甲状旁腺的可行性。前瞻性连续纳入22例患者,使用Firefly系统识别下甲状旁腺,并与倾向性评分匹配的对照组比较。ICG组意外甲状旁腺切除率显著低于对照组(0 vs. 15.9%,P=0.048)。
要点
- ICG联合Firefly系统可术中识别甲状旁腺,平均识别时间203±89秒。
- ICG组意外甲状旁腺切除率0%,对照组15.9%(P=0.048)。
- 荧光持续时间平均20.8±6.0分钟,可用于术中实时定位。
核心结论
- ICG联合近红外光可安全可行地识别BABA机器人甲状腺切除术中的甲状旁腺。
研究设计与样本
- 病例对照研究,前瞻性连续纳入22例BABA机器人甲状腺切除术患者,回顾性选择倾向性评分匹配的对照组。
主要发现
- ICG荧光识别甲状旁腺和甲状腺平均时间分别为203±89秒和207±112秒。
- 甲状旁腺和甲状腺荧光持续时间平均分别为20.8±6.0分钟和20.1±7.3分钟。
- ICG组意外甲状旁腺切除率显著低于对照组(0 vs. 15.9%,P=0.048)。
临床怎么用
- 在BABA机器人甲状腺切除术中可考虑使用ICG联合Firefly系统辅助甲状旁腺定位,以降低意外切除风险。
局限性
- 样本量较小(ICG组22例),对照组为回顾性选择,需更大样本验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: It is unclear whether near-infrared (NIR) light-induced indocyanine green (ICG) fluorescence can effectively identify, and thus permit the preservation of, parathyroid glands in bilateral axillo-breast approach (BABA) robotic thyroidectomy. This case-control study with a prospectively recruited consecutive series and a retrospectively selected control group assessed the usefulness of ICG with Firefly(R) technology to identify the parathyroid glands intraoperatively during BABA robotic thyroidectomy. METHODS: All consecutive patients (N = 22) who were scheduled to undergo BABA robotic thyroidectomy for papillary thyroid carcinoma in December 2013-August 2015 and met the study eligibility criteria were recruited prospectively. ICG fluorescence was used with the Firefly system (NIR illuminator: 805 nm; filter: 825 nm) integrated in the da Vinci Si robot system to identify the lower parathyroid glands. Parathyroid hormone levels were recorded on postoperative days 0, 1, 2, and 14. Propensity score matching was used to identify an age-, gender-, tumor size-, and operation type-matched group of control patients who underwent BABA robotic thyroidectomy without the Firefly system. The two groups were compared in terms of parathyroid-related outcomes. RESULTS: ICG fluorescence-mediated identification of the parathyroid and thyroid glands required on average (range) 203 ± 89 (125-331) and 207 ± 112 (130-356) s, respectively. The mean (range) fluorescence duration in these glands was 20.8 ± 6.0 (16.6-35.8) and 20.1 ± 7.3 (15.5-33.8) min, respectively. The ICG group had a significantly lower rate of incidental parathyroidectomy than the control group (0 vs. 15.9%, P = 0.048). CONCLUSIONS: ICG with NIR light may feasibly and safely identify the parathyroid glands in BABA robotic thyroidectomy.
引用
- Yu HW, Chung JW, Yi JW, et al. Intraoperative localization of the parathyroid glands with indocyanine green and Firefly(R) technology during BABA robotic thyroidectomy. Surg Endosc. 2016. PMID: 27864717. doi:10.1007/s00464-016-5330-y
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。