达芬奇dV5力反馈定量机器人食管裂孔疝修补术中脚张力

Quantification of tension during cruroplasty of robotic-assisted hiatal hernia repairs using force feedback instruments on the da Vinci surgical platform (dV5).
PubMed 文献 · Surg Endosc · 2026-10-12

本研究回顾性分析76例使用达芬奇dV5力反馈系统行机器人食管裂孔疝修补术的患者,采用标准化张力测量方案量化脚关闭张力(CCT)。结果显示平均力为2.92±0.75 N,峰值力为9.60±2.79 N,且峰值力与缺损宽度显著相关(大疝10.38 vs 小疝8.78 N,p=0.011);提出的脚刚度指数(CSI)平均为6.09±2.12 N/cm,与缺损大小无关。

要点

  • dV5力反馈可术中实时量化脚关闭张力,无需外部测量工具。
  • 峰值力随食管裂孔缺损增大而升高,大疝(>2.0 cm)张力更高。
  • 脚刚度指数(CSI)反映脚内在顺应性,可能与初始疝大小无关。

核心结论

  • 使用dV5和标准化张力测量方案可可靠地在体内量化机器人食管裂孔疝修补术中的脚关闭张力。
  • 平均力保持均匀,但峰值力随缺损宽度增大而显著增加。
  • CSI提示脚顺应性可能独立于疝大小,但作为探索性指标需前瞻性验证临床结局。

研究设计与样本

  • 回顾性队列分析,纳入2024年1月至2025年12月单中心单术者使用dV5力反馈系统行机器人食管裂孔疝修补术的76例患者。
  • 采用标准化张力测量方案,对脚施加控制牵引并记录连续力遥测数据。

主要发现

  • 平均食管裂孔缺损宽度为2.16±0.75 cm。
  • 校准测量范围(≤6.5 N)内平均力为2.92±0.75 N,与缺损大小无关(p=0.276)。
  • 平均峰值力为9.60±2.79 N,与缺损宽度显著相关;大疝(>2.0 cm)峰值力高于小疝(10.38 vs. 8.78 N,p=0.011)。
  • 队列平均CSI为6.09±2.12 N/cm,大缺损与小缺损间无显著差异(p=0.888)。

关键数值与推荐

  • 张力测量方案中校准测量范围为≤6.5 N。
  • 大疝定义为缺损宽度>2.0 cm。
  • 主要结局为关闭过程中的平均力、峰值力和CSI。

临床怎么用

  • dV5力反馈可在标准机器人工作流程中实时监测脚关闭张力,无需中断操作。
  • 对较大缺损(>2.0 cm)需注意峰值力升高,可能需调整关闭策略。
  • CSI作为探索性指标尚未指导手术决策,需前瞻性验证。

局限性

  • 回顾性单中心单术者研究,样本量76例,可能存在选择偏倚。
  • CSI为探索性指标,需前瞻性验证 against 临床结局后方可用于指导手术决策。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Tension-free crural closure is critical for durable hiatal hernia repair. Historically, quantifying crural closure tension (CCT) required workflow-disrupting external gauges. The da Vinci 5 (dV5) platform's integrated force feedback allows instantaneous, device-free tension monitoring. This study quantifies in vivo CCT during robotic hiatal hernia repair (RHHR) and introduces an exploratory metric, the Crural Stiffness Index (CSI), defined as peak force per centimeter of crural displacement. METHODS: A retrospective cohort analysis evaluated 76 patients undergoing RHHR at a single institution by a single surgeon using the dV5 system with force feedback instruments between January 2024 and December 2025. Using a standardized tension measurement protocol (TMP), controlled traction was applied to approximate the crura while recording continuous force telemetry. Primary outcomes were mean force, peak force, and CSI during closure. RESULTS: Mean hiatal defect width was 2.16 ± 0.75 cm. Mean force within the calibrated measurement range (≤ 6.5 N) (2.92 ± 0.75 N) remained consistent regardless of defect size (p = 0.276). Peak force averaged 9.60 ± 2.79 N, significantly correlating with defect width; large hernias (> 2.0 cm) required higher tension than small hernias (10.38 vs. 8.78 N, p = 0.011). The cohort's mean CSI was 6.09 ± 2.12 N/cm. CSI did not significantly differ between large and small defects (p = 0.888), suggesting intrinsic compliance is independent of initial hernia size. CONCLUSIONS: Using the dV5 and a standardized TMP, this study reliably quantified, in vivo, CCT during robotic hiatal hernia repairs. While mean forces remained uniform, peak forces significantly increased with larger defect widths. CSI suggested that crural compliance may be independent of hernia size. As a proof of concept, crural tensiometry is feasible within standard robotic workflow. The exploratory CSI requires prospective validation against clinical outcomes before it can guide operative decisions.

来源:Surg Endosc,查看原文。版权归原作者所有。

引用

  1. Ngo K, Langroodi ME, Gatta P, et al. Quantification of tension during cruroplasty of robotic-assisted hiatal hernia repairs using force feedback instruments on the da Vinci surgical platform (dV5). Surg Endosc. 2026. PMID: 42862291. doi:10.1007/s00464-026-13462-5