机器人辅助袖状胃切除术中AI测量工具与标尺测量的一致性研究

Agreement and precision of an AI-based measurement tool during robotic-assisted sleeve gastrectomy.
PubMed 文献 · Surg Endosc · 2026-09-09

该初步研究纳入12例接受机器人辅助袖状胃切除术的成年患者,在胃食管交界处最宽点和角切迹最窄点分别用点对点与轮廓法测量,比较机器人集成AI工具与常规标尺测量的一致性。AI与标尺测量的Pearson相关系数为0.79至0.90,Bland-Altman分析显示各测量偏倚均小于0.1 cm且无比例偏倚证据。

要点

  • AI与标尺测量在活体袖状胃切除术中呈初步一致性
  • 胃食管交界处点对点测量相关性最强(ρ=0.90)
  • 各测量绝对平均差均小于0.1 cm,提示偏倚较低

核心结论

  • 机器人集成AI测量工具与常规标尺测量在活体袖状胃切除术中显示初步一致性,尤其点对点测量。
  • 这些假设生成性发现支持AI辅助测量提升手术精确性与效率的潜力。

研究设计与样本

  • 前瞻性初步研究,纳入12例18岁以上接受机器人辅助袖状胃切除术的成年患者,手术时间2025年8月至2026年1月。
  • 在胃食管交界处最宽点和角切迹最窄点,分别采用点对点与轮廓法测量,比较AI工具与常规标尺测量。
  • 一致性采用Pearson相关分析(含95%CI)、校正BMI的偏相关分析及Bland-Altman分析评估。

主要发现

  • 12例患者均为女性,平均BMI为46.3±6.5 kg/m2。
  • AI与标尺测量的Pearson相关系数在各测量类型中为0.79至0.90。
  • 胃食管交界处点对点测量相关性最强(ρ=0.90;95%CI:0.66-0.97)。
  • 角切迹轮廓测量相关性最低(ρ=0.79;95%CI:0.39-0.94)。
  • Bland-Altman分析显示各测量偏倚低(绝对平均差<0.1 cm),无比例偏倚证据;校正BMI后结果一致。

临床怎么用

  • 结果仅为初步一致性证据,AI测量工具尚不能替代标尺测量用于临床决策。
  • 点对点测量一致性相对更高,可作为后续验证研究的优先评估方式。

局限性

  • 样本量小(n=12)且均为女性,外推性有限。
  • 为初步、假设生成性研究,作者指出需更大规模多中心验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Size and shape of the stomach following a sleeve gastrectomy (SG) appear to impact operative outcomes; however, standardization of these variables is difficult without accurate measurements. Manual measurement techniques are prone to inter-operator variability and can be cumbersome. Artificial Intelligence (AI)-based measurement tools integrated in robotic platforms may provide real-time, precise anatomical assessment. The purpose of this study is to evaluate the preliminary agreement of an AI-based measurement tool relative to conventional ruler measurement during robotic-assisted sleeve gastrectomy (SG) surgery, the most common type of bariatric surgery performed worldwide. METHODS: Adult patients (n = 12, > 18 years) undergoing robotic-assisted SG between August 2025 and January 2026 were included. Measurements (cm) at the widest point of the gastroesophageal (GE) junction and the narrowest point at the incisura were obtained using both point-to-point (P2P) and contour approaches. Data were summarized using mean ± standard deviation (SD). Agreement was assessed using Pearson's correlation analysis with 95% confidence intervals, partial correlation adjusting for BMI, and Bland-Altman analysis. RESULTS: All 12 patients were female with a mean BMI of 46.3 ± 6.5 kg/m2. Pearson correlation coefficients between AI and ruler measurements ranged from 0.79 to 0.90 across all measurement types. P2P measurements at the GE junction demonstrated the strongest correlation (ρ = 0.90; 95% CI: 0.66-0.97), while contour measurements at the incisura showed the lowest (ρ = 0.79; 95% CI: 0.39-0.94). Bland-Altman analysis demonstrated low bias across all measurements (absolute mean differences < 0.1 cm) with no evidence of proportional bias. Partial correlation analysis adjusting for BMI showed consistent results. CONCLUSION: This pilot study demonstrates preliminary agreement between commercially available robotic-integrated AI measurement tool and standard ruler measurement during live SG, particularly for P2P measurements. These hypothesis-generating findings support the potential of AI-assisted measurement to enhance surgical precision and efficiency, providing a foundation for future larger-scale, multicenter validation studies.