PubMed 文献 · Surg Endosc · 2026-05-28

本研究回顾性分析单术者连续13例机器人直肠癌切除术,从系统日志提取Cadiere钳(静态牵拉)和开窗双极钳(动态牵拉)的力数据,比较力反馈关闭、低、中三种设置。结果显示静态牵拉平均力随灵敏度升高显著降低(P=0.039),动态牵拉峰值力随设置升高显著被抑制(P=0.033),无组织创伤相关术中并发症。

要点

  • 力反馈在静态牵拉中显著降低平均力,灵敏度越高力越小。
  • 力反馈在动态操作中不改变平均力,但显著抑制峰值力。
  • 力反馈可调器械力,静态减持续应力,动态限峰值力。

核心结论

  • 达芬奇5力反馈技术根据器械功能角色有效调节手术力,静态牵拉时显著降低持续应力,动态操作时限制峰值力,可能提升机器人直肠手术安全性。

研究设计与样本

  • 回顾性分析单术者连续13例机器人直肠癌切除术,从系统日志提取Cadiere钳和开窗双极钳的力数据。
  • 比较力反馈关闭(N=3)、低(N=11)、中(N=5)三种设置,低和中设置病例因术中调整而重叠。
  • 计算每种设置的累积力使用时间,采用Kruskal-Wallis检验评估统计学显著性。

主要发现

  • Cadiere钳(静态牵拉)分析91,000+秒,平均力随力反馈灵敏度升高显著降低:关闭3.07 N,低2.58 N,中2.03 N,P=0.039。
  • 开窗双极钳(动态牵拉)分析68,000+秒,平均力无显著差异,但中位最大(峰值)力随力反馈设置升高显著被抑制:关闭36.19 N,低18.82 N,中10.06 N,P=0.033。
  • 无组织创伤相关术中并发症发生。

临床怎么用

  • 静态牵拉可提高力反馈灵敏度以降低持续应力。
  • 动态操作可提高力反馈设置以限制峰值力,起「安全刹车」作用。

局限性

  • 回顾性单术者小样本先导研究,力反馈低和中设置病例因术中调整而重叠。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The da Vinci 5 (DV5) introduces force feedback (FFb) technology to robotic surgery, aiming to enhance safety through haptic sensation. However, its quantitative impact on specific tissue manipulations in rectal cancer surgery remains unclear. This study evaluated the clinical significance of FFb by analyzing force metrics across different instrument roles. METHODS: We retrospectively analyzed 13 consecutive robotic rectal cancer resections performed by a single expert surgeon. Force data were extracted from system logs for the Cadiere Forceps (static retraction) and Fenestrated Bipolar Forceps (dynamic retraction). Comparisons were made between three settings: Off (N = 3), Low (N = 11), and Medium (N = 5). For "Low" and "Medium" settings, cases overlapped as settings were adjusted intraoperatively. Cumulative force usage time was calculated for each setting. Statistical significance was assessed using the Kruskal-Wallis test. RESULTS: A total of 91,000 + seconds for Cadiere and 68,000 + seconds for Fenestrated Bipolar forceps were analyzed. For the Cadiere Forceps (static retraction), the mean force was significantly reduced as FFb sensitivity increased (Off: 3.07 N, Low: 2.58 N, Medium: 2.03 N; P = 0.039). For the Fenestrated Bipolar Forceps (dynamic retraction), while the mean force showed no significant difference, the median maximum (peak) force was significantly suppressed with higher FFb settings (Off: 36.19 N, Low: 18.82 N, Medium: 10.06 N; P = 0.033). No intraoperative complications related to tissue trauma occurred. CONCLUSIONS: FFb technology in the DV5 effectively modulates surgical force based on the functional role of the instrument. It significantly reduces sustained stress during static retraction and serves as a "safety brake" to cap peak forces during dynamic maneuvers, potentially enhancing the safety of robotic rectal surgery.

来源:Surg Endosc,查看原文。版权归原作者所有。

引用

  1. Nishi Y, Hirano Y, Ishiyama Y, et al. Quantitative analysis of da Vinci 5 force feedback technology in robotic rectal surgery: a preliminary pilot study. Surg Endosc. 2026. PMID: 42209862. doi:10.1007/s00464-026-12842-1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。