文献速递胃肠外科围术期管理与 ERAS

荧光引导袖状胃切除与胃旁路术:新型胃校准管首次人体试验

Fluorescence-guided sleeve gastrectomy and gastric bypass: first-in-human trial using novel gastric calibration tube.
PubMed 文献 · Surg Endosc · 2025-09-02

本研究在IRB批准下由两名外科医生使用Endolumik荧光引导校准管(EGCT)完成21例腹腔镜袖状胃切除术(SG)和10例胃旁路术(GB)。无器械相关不良事件,手术团队对器械体验评价良好,EGCT改善了袖状胃和胃小囊构建时的可视化并提高GB中器械置入信心,但需进一步研究临床结局影响。

要点

  • EGCT在SG和GB中无器械相关不良事件。
  • 手术团队评价EGCT体验良好或非常好,推荐意愿平均9/10。
  • EGCT改善袖状胃和胃小囊构建可视化并提高GB置入信心。

核心结论

  • EGCT改善了袖状胃和胃小囊构建时的可视化,并提高GB中器械置入信心,外科医生愿意推荐该器械。
  • 需进一步研究评估对临床结局的影响。

研究设计与样本

  • 首次人体试验,经IRB批准(NCT05486325),两名外科医生使用EGCT完成21例腹腔镜SG和10例GB。
  • 手术和麻醉人员填写问卷评价EGCT使用体验。

主要发现

  • 无器械相关不良事件。
  • EGCT由SG/GB中10/5名独特手术团队成员和20/5名麻醉人员使用。
  • 所有手术团队成员评价器械体验良好或非常好(对比标准探条)。
  • 袖状胃构建时可视化评价良好或非常好,胃小囊构建时评价非常好。
  • 所有手术成员对器械实现一致袖状胃或胃小囊 sizing 有信心。

临床怎么用

  • EGCT可辅助腹腔镜SG和GB中胃校准和漏测试,改善可视化并提高操作信心。

局限性

  • 需进一步研究评估对临床结局的影响。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: Fluorescence-guided surgery employs a near-infrared emitting dye or light source to enhance intraoperative visualization. This study reports the first-in-human application of the Endolumik fluorescence-guided calibration tube (EGCT) during laparoscopic sleeve gastrectomy (SG) and gastric bypass (GB). METHODS: Under IRB approval (NCT05486325), two surgeons performed 21 laparoscopic SG and 10 GB procedures using EGCT. Evacuation of gastric contents, gastric sleeve/pouch calibration, and leak testing were performed. Surgical and anesthesia providers completed surveys evaluating their EGCT experience. RESULTS: No device-related adverse events occurred. EGCT was used by 10/5 unique surgical team members during SG/GB, respectively, and by 20/5 anesthesia providers. All surgical team members rated device experience as good or very good compared to the standard bougie. Visualization during gastric sleeve construction was rated as good or very good, and very good for gastric pouch construction. All surgical members reported confidence in the device enabling consistent gastric sleeve or pouch sizing. EGCT increased surgeon confidence level for device introduction during GB. Among anesthesia providers, 84% rated their experience during SG with the device as good or very good, including 89.5% of CRNAs. Additionally, 75%/90% of SG/GB clinicians believed the device reduced the likelihood of adverse events compared to standard devices. The average likelihood to recommend the EGCT was 9 out of 10 among surgeons. CONCLUSION: EGCT improved visualization during gastric sleeve and pouch construction and increased confidence during device introduction for GB; surgeons would recommend the device to colleagues. Further research is needed to assess the impact on clinical outcomes.