磁括约肌增强装置失效模式:单中心病例系列与文献综述
BACKGROUND: Magnetic sphincter augmentation (MSA) is an effective surgical treatment for GERD. Removal rates are reported around 4-6%, with dysphagia being the most common indication for removal.…
要点
- MSA因吞咽困难或反流复发取出者中,多数存在装置滑移和/或食管裂孔疝。
- 文献报道MSA取出率0-12.6%(中位4.7%),随访时间较短。
- 装置滑移/移位可能被低估,改进置入技术或可提高长期成功率。
核心结论
- 长期随访中,因吞咽困难或反流复发取出的MSA患者呈现装置滑移或移位模式,与360度胃底折叠术失效模式相似。
- 装置滑移或移位作为症状病因可能被低估。
研究设计与样本
- 单中心回顾性研究,纳入2014至2024年接受MSA置入并取出的患者,共42例置入,其中12例在本中心置入、1例在外院置入后取出。
- 结合影像学、内镜和手术记录评估术前术后食管裂孔疝或装置位置异常,并系统综述MSA取出文献。
主要发现
- 取出原因为吞咽困难和/或反流复发,中位随访41.1个月。
- 文献综述显示取出率0-12.6%(中位4.7%),随访时间较短,最常见取出原因为吞咽困难,其次为反流复发或持续。
- 文献中装置移位/食管裂孔疝罕见。
临床怎么用
- 对MSA术后出现吞咽困难或反流复发的患者,应评估装置滑移或移位及食管裂孔疝。
- 关注置入技术改进、患者选择及生活方式指导以提升长期成功率。
局限性
- 单中心回顾性设计,样本量小(42例置入,13例取出)。
- 文献综述中随访时间较短,可能低估装置移位/裂孔疝发生率。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Magnetic sphincter augmentation (MSA) is an effective surgical treatment for GERD. Removal rates are reported around 4-6%, with dysphagia being the most common indication for removal. This rate has increased over time, and the pattern of device failure has not been well established. This study characterizes a consistent pattern of MSA slippage or migration and contextualizes this within the existing literature on MSA failure. METHODS: This is a single-institution retrospective review of patients who underwent MSA placement and device removal between 2014 and 2024. Radiographic and endoscopic images and operative reports were reviewed for the presence of pre-operative and post-operative hiatal hernia or device malposition. We also conducted a comprehensive review of existing literature on MSA explantation, with particular attention to device slippage, malposition, and hiatal hernia. RESULTS: 42 patients underwent MSA placement at our tertiary academic institution. Twelve of these, plus one patient who had MSA placed elsewhere, underwent device removal for symptoms of dysphagia and/or recurrent reflux (28%) with a median follow-up of 41.1 months. Ten of these 13 patients showed evidence of MSA slippage and/or hiatal hernia. In comparison, our literature review revealed explant rates of 0-12.6% (median 4.7%) with overall shorter terms of follow-up. Similarly, the most common reason for explantation was dysphagia, followed by recurrent or persistent reflux. However, device migration/hiatal herniation was a rare finding. CONCLUSION: In the long-term follow-up of MSA patients with dysphagia or recurrent reflux, we observe a pattern of device slippage or migration. This pattern mirrors the failure pattern of the 360-degree fundoplication. We suspect an under-appreciation of device slippage or migration as the etiology for these symptoms. While MSA is effective, continued improvements on implantation technique, coupled with careful patient selection and lifestyle counseling, may increase its long-term success rate.
引用
- Bloomsburg SJ, Duncan AJ, Chen S, et al. The failure pattern for the magnetic sphincter augmentation device: a single-institution case series with literature review. Surg Endosc. 2025. PMID: 40571793. doi:10.1007/s00464-025-11842-x
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。