荧光成像在甲状腺切除术中可视化喉返神经:65例81条神经分析
这项前瞻性研究纳入65例甲状腺切除术患者(16例双侧切除),用自发荧光成像检测喉返神经,并与白光对比。结果显示荧光下喉返神经显示更清晰,灵敏度和特异度均为100%,可视化神经长度是白光的2.5倍。
要点
- 荧光成像对喉返神经检测的灵敏度和特异度均为100%。
- 荧光下可见喉返神经长度是白光的2.5倍(平均1.90 cm vs 0.76 cm)。
- 荧光下30条神经和45条分支均清晰可见,白光下分别为17条和22条。
核心结论
- 在65例患者和81条神经中,甲状腺切除术中自发荧光神经成像显著且一致地改善了喉返神经的检测,灵敏度和特异度均为100%。
研究设计与样本
- 前瞻性研究,纳入2022年1月至4月行甲状腺切除术的65例患者(16例双侧甲状腺切除),使用白光和荧光两种模式。
主要发现
- 所有81条观察到的喉返神经在荧光下显示更清晰(平均强度134.3 RFU),高于甲状腺(平均33.7,p < 0.001)和背景组织(平均14.4,p < 0.001)。
- 森林图显示喉返神经强度与其他组织无重叠。
- 荧光下30条喉返神经和45条神经分支均清晰可见,白光下分别为17条和22条(均p < 0.001)。
- 荧光下可见神经长度是白光的2.5倍(平均1.90 cm vs 0.76 cm,p < 0.001)。
关键数值与推荐
- 喉返神经检测的灵敏度和特异度均为100%。
- 荧光下喉返神经平均强度134.3 RFU,甲状腺平均33.7 RFU(p < 0.001),背景组织平均14.4 RFU(p < 0.001)。
- 荧光下可见神经长度平均1.90 cm,白光下平均0.76 cm(p < 0.001)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Recurrent laryngeal nerve (RLN) injury after thyroidectomy is relatively common. Locating the RLN prior to thyroid dissection is paramount to avoid injury. We developed a fluorescence imaging system that permits nerve autofluorescence. We aimed to determine the sensitivity and specificity of fluorescence imaging at detecting the RLN relative to thyroid and other background tissue and compared it to white light. METHODS: In this prospective study, 65 patients underwent thyroidectomy from January to April 2022 (16 bilateral thyroid resections) using white and fluorescent light. Fluorescence intensity [relative fluorescence units (RFU)] was recorded for RLN, thyroid, and background. RFU mean, minimum, and maximum values were calculated using Image J software. Thirty randomly selected pairs of white and fluorescent light images were independently reviewed by two examiners to compare RLN detection rate, number of branches, and length and minimum width of nerves visualized. Parametric and nonparametric statistical analysis was performed. RESULTS: All 81 RNLs observed were visualized more clearly under fluorescence (mean intensity, µ = 134.3 RFU) than either thyroid (µ = 33.7, p < 0.001) or background (µ = 14.4, p < 0.001). Forest plots revealed no overlap between RLN intensity and that of either other tissue. Sensitivity and specificity for RLN were 100%. All 30 RLNs and all 45 nerve branches were clearly visualized under fluorescence, versus 17 and 22, respectively, with white light (both p < 0.001). Visible nerve length was 2.5 × as great with fluorescence as with white light (µ = 1.90 vs. 0.76 cm, p < 0.001). CONCLUSIONS: In 65 patients and 81 nerves, RLN detection was markedly and consistently enhanced with autofluorescence neuro-imaging during thyroidectomy, with 100% sensitivity and specificity.
引用
- Dip F, Falco J, White K, et al. Fluorescence imaging to visualize the recurrent laryngeal nerve during thyroidectomy procedures: analysis of 65 cases and 81 nerves. Surg Endosc. 2024. PMID: 38168731. doi:10.1007/s00464-023-10627-4
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。