腹腔镜手术中气体泄漏的特征与缓解:一项国际前瞻性多中心队列临床试验
该研究在三个中心对40例择期腹腔镜手术患者(胆囊切除术36例、疝修补术4例)使用移动光学气体成像仪前瞻性检测穿刺器相关气体泄漏,并评估真空环装置LeakTrap™的缓解效果。共检查1643次穿刺器置入操作,对照组819次、干预组824次。对照组中60%的泄漏为3类(高能湍流射流),干预组仅8.7%(3类 vs 非3类,P<0.0001)。LeakTrap在手术可接受性方面表现良好,未造成显著工作流程中断。
要点
- 腹腔镜穿刺器相关气体泄漏可通过光学气体成像灵敏检测并分级。
- LeakTrap真空环装置可有效减少高能3类气体泄漏(8.7% vs 60%,P<0.0001)。
- 该装置手术可接受性好,未显著干扰手术流程。
核心结论
- 腹腔镜气体泄漏可被灵敏检测,并使用现成技术持续有效缓解,对最常见的高能器械交换泄漏影响最大。
研究设计与样本
- 国际前瞻性多中心队列临床试验,在三个医院纳入40例择期腹腔镜手术患者(平均年龄58.6岁,男性20例),其中胆囊切除术36例、疝修补术4例,分为对照组20例和干预组20例。
主要发现
- 共检查1643次穿刺器器械置入操作,对照组819次(平均每例41次),干预组824次(平均每例41.2次)。
- 在948次(62.6%)可视穿刺器器械置入操作中检测到气体泄漏,另有129次(7.8%)成像被遮挡。
- 对照组泄漏中0类和3类分别占14.8%(110/742)和60%(445/742),干预组分别占59.1%(456/770)和8.7%(67/772)(3类 vs 非3类,P<0.0001,χ2检验)。
- LeakTrap在手术可接受性方面表现良好,未造成显著工作流程中断。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Gas leaks polluting the operating room are common in laparoscopy. Studies defining methods for sensitive leak characterisation and mechanical mitigation in real world settings are, however, lacking. METHODS: Mobile optical gas imagers (both a miniaturised Schlieren system and sensitive tripod-mounted near-infrared carbon dioxide camera (GF343, FLIR)) prospectively defined trocar-related gas leaks occurring either spontaneously or with instrumentation during planned laparoscopic surgery at three hospitals. A boutique Matlab-based analyser using sequential frame subtraction categorised leaks (class 0-no observable leak; class 1-marginally detectable leak; class 2-short-lived plume; class 3-energetic, turbulent jet). Concurrently, the usefulness of a novel vacuum-ring device (LeakTrap™, Palliare, Ireland) designed as a universal adjunct for existing standard laparoscopic ports at both abdominal wall and port valve level was determined similarly in a phase I/11 clinical trial along with the device's useability through procedural observation and surgeon questionnaire. RESULTS: With ethical and regulatory approval, 40 typical patients (mean age 58.6 years, 20 males) undergoing planned laparoscopic cholecystectomy (n = 36) and hernia repair (n = 4) were studied comprising both control (n = 20) and intervention (n = 20) cohorts. Dual optical gas imaging was successfully performed across all procedures with minimal impact on procedural flow. In total, 1643 trocar instrumentations were examined, 819 in the control group (mean 41 trocar instrumentations/procedure) and 824 in the intervention group (mean 41.2 trocar instrumentations/procedure). Gas leaks were detected during 948(62.6%) visualised trocar instrumentations (in 129-7.8%-the imaging was obscured). 14.8% (110/742) and 60% (445/742) of leaks in control patients were class 0 and 3, respectively, versus 59.1% (456/770) and 8.7% (67/772) in the interventional group (class 3 v non-class 3, p < 0.0001, χ2). The Leaktrap proved surgically acceptable without significant workflow disruption. CONCLUSION: Laparoscopic gas leaks can be sensitively detected and consistently, effectively mitigated using straightforward available-now technology with most impact on the commonest, highest energy instrument exchange leaks.
引用
- Nolan K, Khan MF, Riva P, et al. Characterisation and mitigation of gas leaks at laparoscopy: an international prospective, multi-center cohort clinical trial. Surg Endosc. 2023. PMID: 37985488. doi:10.1007/s00464-023-10536-6
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。