经环氧乙烷灭菌的疝外科补片在英国法规下采用的评估

Evaluation of hernia surgical meshes sterilised with ethylene oxide for adoption under UK regulations.
PubMed 文献 · Surg Endosc · 2023-09-20

本研究对经环氧乙烷灭菌的英国低成本补片进行力学性能与体外生物相容性表征,单轴拉伸显示LCM1断裂应力显著高于LCM2而杨氏模量显著低于LCM2,断裂应变无显著差异;人真皮成纤维细胞在LCM2上的增殖优于LCM1且终末时间点与对照相似,环氧乙烷灭菌影响LCM1的细胞附着但不影响LCM2的体外生物相容性。

要点

  • 环氧乙烷灭菌影响低成本补片力学性能,但仍较商用补片更接近原生组织。
  • LCM1断裂应力显著高于LCM2(p<0.0001),杨氏模量显著低于LCM2(p<0.05)。
  • LCM2体外生物相容性未受环氧乙烷灭菌影响,而LCM1的细胞附着受影响。

核心结论

  • 环氧乙烷灭菌影响低成本补片的力学性能,但其数值仍较商用补片更接近原生组织。
  • LCM2的体外生物相容性未受环氧乙烷灭菌影响,人真皮成纤维细胞可在补片上附着和增殖,而环氧乙烷灭菌影响LCM1的细胞附着。
  • 需要对全球卫生系统潜在节省进行更详细的成本分析,以强化这一节俭创新的成本效益。

研究设计与样本

  • 研究对经环氧乙烷灭菌的英国低成本补片进行表征,采用单轴拉伸试验测量力学性能。
  • 通过在人真皮成纤维细胞条件培养基中培养检测细胞活力和形态,并用alamarBlue法计算附着于补片上人真皮成纤维细胞的代谢活性和增殖来评估生物相容性。

主要发现

  • LCM1的断裂应力显著高于LCM2(p<0.0001)。
  • LCM1的杨氏模量显著低于LCM2(p<0.05),断裂应变无显著差异。
  • 细胞活力和形态在低成本补片与对照之间无显著差异。
  • 人真皮成纤维细胞在LCM2上的增殖优于LCM1,终末时间点数值与对照相似。

局限性

  • 需要对全球卫生系统潜在节省进行更详细的成本分析,以强化这一节俭创新的成本效益。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Low-cost meshes (LCM) have been successfully used in low-income countries (LIC) over the past decades, demonstrating comparable surgical outcomes to commercial meshes at a fraction of the cost. However, LIC sterilisation standards (autoclave sterilisation at 121 °C) do not meet UK regulations for medical devices, which require either ethylene oxide (EO) sterilisation or steam sterilisation at 134 °C. Therefore, the aim of this study was to sterilise UK LCM and characterise their mechanical properties and in vitro biocompatibility to verify whether EO sterilisation causes changes in the mechanical properties and biocompatibility of LCM. METHODS: EO sterilised LCM were used. Uniaxial tensile tests were performed to measure mechanical properties. Biocompatibility was measured through viability and morphology of Human Dermal Fibroblasts (HDFs) cultured in mesh-conditioned media, and by calculating the metabolic activity and proliferation of HDFs attached on the meshes, with alamarBlue assay. RESULTS: Break stress of LCM1 was significantly higher than LCM2 (p < 0.0001), while Young's modulus of LCM1 was significantly lower than LCM2 (p < 0.05) and there was no significant difference in break strain. Viability and morphology showed no significant difference between LCM and control. Attachment and proliferation of HDFs on LCM showed a better proliferation on LCM2 than LCM1, with values similar to the control at the final time point. CONCLUSIONS: We demonstrated that EO sterilisation affects LCM mechanical properties, but they still have values closer to the native tissues than the commercially available ones. We also showed that in vitro biocompatibility of LCM2 is not affected by EO sterilisation, as HDFs attached and proliferated on the mesh, while EO affected attachment on LCM1. A more detailed cost analysis of the potential savings for healthcare systems around the world needs to be performed to strengthen the cost-effectiveness of this frugal innovation.