长时间器械操作对穿刺器气体泄漏的影响

The influence of prolonged instrument manipulation on gas leakage through trocars.
PubMed 文献 · Surg Endosc · 2023-07-13

结果显示操作前后静态和动态泄漏率无大差异,泄漏率范围为0.0至5.58 L/min,但长时间操作导致两个穿刺器可见损坏并暴露其他穿刺器的非预期泄漏通路。

要点

  • 长时间器械操作未增加穿刺器气体泄漏率。
  • 穿刺器泄漏率范围为0.0至5.58 L/min。
  • 部分穿刺器因长时间操作出现损坏或暴露生产缺陷相关泄漏。

核心结论

  • 长时间器械操作未增加穿刺器气体泄漏率,尽管部分穿刺器出现损坏。
  • 穿刺器气体泄漏确实存在,由不同穿刺器特异性机制和设计问题引起。

研究设计与样本

  • 建立体外模型,穿刺器安装于充入CO2至15 mmHg的密闭容器,使用线性平台进行长时间器械操作并施加径向固定负载模拟腹壁反作用力。
  • 在器械操作前、后及操作过程中测量气体泄漏。

主要发现

  • 器械操作后每个穿刺器的泄漏率范围为0.0至5.58 L/min。
  • 长时间操作前后静态和动态测量的泄漏率无大差异。
  • 长时间器械操作导致两个穿刺器可见损坏,并在其他穿刺器中发现可能与生产缺陷相关的非预期泄漏通路。

临床怎么用

  • 普通外科医生在腹腔镜手术中应知晓穿刺器存在气体泄漏风险,泄漏与穿刺器特异性机制和设计问题相关。
  • 长时间器械操作本身不增加泄漏率,但需关注个别穿刺器损坏及生产缺陷导致的非预期泄漏。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: During laparoscopic surgery, CO2 insufflation gas could leak from the intra-abdominal cavity into the operating theater. Medical staff could therefore be exposed to hazardous substances present in leaked gas. Although previous studies have shown that leakage through trocars is a contributing factor, trocar performance over longer periods remains unclear. This study investigates the influence of prolonged instrument manipulation on gas leakage through trocars. METHODS: Twenty-five trocars with diameters ranging from 10 to 15 mm were included in the study. An experimental model was developed to facilitate instrument manipulation in a trocar under loading. The trocar was mounted to a custom airtight container insufflated with CO2 to a pressure of 15 mmHg, similar to clinical practice. A linear stage was used for prolonged instrument manipulation. At the same time, a fixed load was applied radially to the trocar cannula to mimic the reaction force of the abdominal wall. Gas leakage was measured before, after, and during instrument manipulation. RESULTS: After instrument manipulation, leakage rates per trocar varied between 0.0 and 5.58 L/min. No large differences were found between leakage rates before and after prolonged manipulation in static and dynamic measurements. However, the prolonged instrument manipulation did cause visible damage to two trocars and revealed unintended leakage pathways in others that can be related to production flaws. CONCLUSION: Prolonged instrument manipulation did not increase gas leakage rates through trocars, despite damage to some individual trocars. Nevertheless, gas leakage through trocars occurs and is caused by different trocar-specific mechanisms and design issues.

来源:Surg Endosc,查看原文。版权归原作者所有。

引用

  1. Robertson D, van Duijn M, Arezzo A, et al. The influence of prolonged instrument manipulation on gas leakage through trocars. Surg Endosc. 2023. PMID: 37442835. doi:10.1007/s00464-023-10240-5