压力限制器降低内镜注气腔内压的临床前盲法试验
这项临床前盲法随机对照试验纳入79名内镜医师,比较食管ESD中自动注气系统有无压力限制器(防止腔内压超过25 mmHg)的效果。有压力限制器组腔内压显著低于无限制器组(10.6 ± 4.61 vs. 16.25 ± 7.51 mmHg,p < 0.05),且无动物死亡,而无限制器组有2头猪死于张力性气胸。内镜医师对设备可用性的主观评价无显著差异。
要点
- 压力限制器组食管ESD腔内压显著低于无限制器组(10.6 ± 4.61 vs. 16.25 ± 7.51 mmHg,p < 0.05)。
- 无限制器组2头猪死于张力性气胸,压力限制器组无死亡。
- 压力限制器组管腔扩张、吸引容易度和视野可重复性评价较差但均在可接受范围内。
局限性
- 原文未述局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Optimal visualization and safety have always been essential in performing any type of endoscopic surgery. However, the safety of automatic gastrointestinal (GI) insufflation has yet to be thoroughly studied, especially when combined with manual insufflation. The current study aimed to verify whether the pressure limiter could lower GI endoluminal pressure during endoscopic procedures and affect the behavioral patterns of endoscopists. METHODS: A preclinical blinded trial was conducted on endoscopists who had no knowledge regarding the presence of the pressure limiter that prevents a GI endoluminal pressure above 25 mmHg. Endoscopists in group A performed esophageal endoscopic submucosal dissection (ESD) with our insufflation device equipped with the pressure limiter, whereas those in group B performed the same procedure without the pressure limiter. During all procedures, endoluminal pressure was continuously monitored. The primary endpoint of the current study was to measure the endoluminal pressure with or without the pressure limiter during esophageal ESD, while the secondary endpoint was to evaluate the effect of the pressure limiter on intraesophageal pressure and perioperative outcomes during esophageal ESD. A questionnaire survey was conducted after each session. RESULTS: A total of 79 endoscopists were included in this randomized control study. Group A had significantly lower endoluminal pressure than group B (10.6 ± 4.61 vs. 16.25 ± 7.51 mmHg, respectively; p < 0.05). Although two pigs in group B died from tension pneumothorax, none in group A died. Evaluation of lumen expansion, ease of aspiration, and visual field reproducibility were poorer in group A than in group B, although all fell within the acceptable range. Subjective evaluation of usability was divided into two categories, Excellent/Good and Poor/Bad, with no significant differences in any of the items. CONCLUSIONS: This preclinical study showed that endoscopic treatment with an automatic insufflation system could be performed at lower endoluminal pressure with a pressure limiter, which had no adverse effects on the endoscopist's feels on endoscopic procedures with the device.
引用
- Ushimaru Y, Takahashi T, Yamashita K, et al. Translation from manual to automatic endoscopic insufflation enhanced by a pressure limiter. Surg Endosc. 2022. PMID: 35041055. doi:10.1007/s00464-022-09040-0
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。