耳蜗电图引导的机器人辅助人工耳蜗植入系统开发与表征
本研究开发并表征了一种耳蜗电图(ECochG)引导的机器人辅助人工耳蜗植入系统,旨在监测并适应实时ECochG信号,实现受控且一致的电极阵列插入。台架概念验证实验评估了系统检测模拟ECochG信号变化并机器人自适应插入的能力;在活体羊预实验中表征了机器人辅助插入期间ECochG记录的信噪比和幅度。系统具有微伏级信号分辨率和信号变化检测后<100毫秒的响应时间,动物实验证明能记录高信噪比和足够幅度的ECochG信号。
要点
- 系统可检测电极插入中ECochG信号实时下降并立即调整插入动作。
- 系统响应时间<100毫秒,快于人类反应。
- 活体羊实验显示系统能记录高信噪比和足够幅度的ECochG信号。
核心结论
- ECochG引导的机器人辅助人工耳蜗植入系统可检测电极阵列插入过程中ECochG信号的实时下降并立即改变插入运动。
- 该系统可能为外科医生提供监测和减少人工耳蜗植入相关创伤的手段,超越手动插入技术以改善听力结果。
研究设计与样本
- 实验研究,包括概念验证台架实验和活体羊预实验。
- 系统由电极阵列插入驱动单元、耳蜗外记录电极模块和控制台组成。
主要发现
- 系统具有微伏级信号分辨率和信号变化检测后<100毫秒的响应时间。
- 动物实验结果表明系统能够记录高信噪比和足够幅度ECochG信号。
临床怎么用
- 该系统可能帮助外科医生在人工耳蜗植入术中实时监测ECochG信号并调整插入动作以减少创伤。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Electrocochleography (ECochG) is increasingly being used during cochlear implant (CI) surgery to detect and mitigate insertion-related intracochlear trauma, where a drop in ECochG signal has been shown to correlate with a decline in hearing outcomes. In this study, an ECochG-guided robotics-assisted CI insertion system was developed and characterized that provides controlled and consistent electrode array insertions while monitoring and adapting to real-time ECochG signals. STUDY DESIGN: Experimental research. SETTING: A research laboratory and animal testing facility. METHODS: A proof-of-concept benchtop study evaluated the ability of the system to detect simulated ECochG signal changes and robotically adapt the insertion. Additionally, the ECochG-guided insertion system was evaluated in a pilot in vivo sheep study to characterize the signal-to-noise ratio and amplitude of ECochG recordings during robotics-assisted insertions. The system comprises an electrode array insertion drive unit, an extracochlear recording electrode module, and a control console that interfaces with both components and the surgeon. RESULTS: The system exhibited a microvolt signal resolution and a response time <100 milliseconds after signal change detection, indicating that the system can detect changes and respond faster than a human. Additionally, animal results demonstrated that the system was capable of recording ECochG signals with a high signal-to-noise ratio and sufficient amplitude. CONCLUSION: An ECochG-guided robotics-assisted CI insertion system can detect real-time drops in ECochG signals during electrode array insertions and immediately alter the insertion motion. The system may provide a surgeon the means to monitor and reduce CI insertion-related trauma beyond manual insertion techniques for improved CI hearing outcomes.
引用
- Henslee AM, Kaufmann CR, Andrick MD, et al. Development and Characterization of an Electrocochleography-Guided Robotics-Assisted Cochlear Implant Array Insertion System. Otolaryngol Head Neck Surg. 2021. PMID: 34609909. doi:10.1177/01945998211049210
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。