文献速递耳科与耳神经

临床颞骨CT全自动测量耳蜗导管长度

Fully Automated Measurement of Cochlear Duct Length From Clinical Temporal Bone Computed Tomography.
PubMed 文献 · Laryngoscope · 2021-09-18

本研究提出并验证了一种全自动测量耳蜗尺寸(包括耳蜗导管长度CDL)的新方法,结合深度学习分割与几何分析。在159例测试CT中,98.1%的扫描成功获得测量,自动与手动测量的R²为A值0.87、B值0.70、CDL0.71,且预测与实际的阵列尖端位置差异为0.57 mm。

要点

  • 全自动方法在98.1%的临床CT中成功测量耳蜗尺寸。
  • 自动与手动测量一致性:A值R²=0.87,B值R²=0.70,CDL R²=0.71。
  • 对反对耳蜗轴阵列,预测与影像尖端位置差0.57 mm。

核心结论

  • 全自动方法可基于临床CT完成耳蜗分析,无需放射学经验且无用户相关变异。

研究设计与样本

  • 横断面研究,使用165例手动分割CT训练深度学习算法,159例测试CT进行验证。
  • 另纳入27例耳蜗植入者的术前和术后扫描比较预测与实际阵列插入深度。

主要发现

  • 测量成功率为98.1%。
  • CDL至900°均值35.52 mm(SD 2.06,范围30.91-40.50)。
  • A值均值8.88 mm(SD 0.47,范围7.67-10.49),B值均值6.38 mm(SD 0.42,范围5.16-7.38)。
  • 自动与手动测量的R²:A值0.87,B值0.70,CDL 0.71。
  • 对反对耳蜗轴阵列,影像与预测阵列尖端位置距离0.57 mm(SD 1.25,范围0.13-5.28)。

临床怎么用

  • 可用于临床CT的耳蜗尺寸全自动测量,辅助耳蜗植入术前评估。
  • 无需放射学经验,减少用户间变异。

局限性

  • 证据等级为3级。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES/HYPOTHESIS: To present and validate a novel fully automated method to measure cochlear dimensions, including cochlear duct length (CDL). STUDY DESIGN: Cross-sectional study. METHODS: The computational method combined 1) a deep learning (DL) algorithm to segment the cochlea and otic capsule and 2) geometric analysis to measure anti-modiolar distances from the round window to the apex. The algorithm was trained using 165 manually segmented clinical computed tomography (CT). A Testing group of 159 CTs were then measured for cochlear diameter and width (A- and B-values) and CDL using the automated system and compared against manual measurements. The results were also compared with existing approaches and historical data. In addition, pre- and post-implantation scans from 27 cochlear implant recipients were studied to compare predicted versus actual array insertion depth. RESULTS: Measurements were successfully obtained in 98.1% of scans. The mean CDL to 900° was 35.52 mm (SD, 2.06; range, [30.91-40.50]), the mean A-value was 8.88 mm (0.47; [7.67-10.49]), and mean B-value was 6.38 mm (0.42; [5.16-7.38]). The R2 fit of the automated to manual measurements was 0.87 for A-value, 0.70 for B-value, and 0.71 for CDL. For anti-modiolar arrays, the distance between the imaged and predicted array tip location was 0.57 mm (1.25; [0.13-5.28]). CONCLUSION: Our method provides a fully automated means of cochlear analysis from clinical CTs. The distribution of CDL, dimensions, and cochlear quadrant lengths is similar to those from historical data. This approach requires no radiographic experience and is free from user-related variation. LEVEL OF EVIDENCE: 3 Laryngoscope, 132:449-458, 2022.