文献速递耳科与耳神经

人工耳蜗使用者皮层听觉诱发电位的比较分析

Comparative Analysis of Cortical Auditory Evoked Potential in Cochlear Implant Users.
PubMed 文献 · Ear Hear · 2021-01-01

结果显示两者P1-N1-P2波形清晰可靠且相关性良好;刺激水平降低时潜伏期延长、波幅下降。eCAEP因直接刺激、干扰小且设置灵活,可能更适于在皮层水平验证言语检测,可作为人工耳蜗调试的替代方法。

要点

  • eCAEP与aCAEP的P1-N1-P2波形均清晰可靠且相关性良好。
  • 刺激水平从MCL降至50% MCL时潜伏期延长、波幅下降。
  • eCAEP直接刺激干扰小、设置灵活,或更适于皮层水平言语检测验证。

核心结论

  • 本研究成功在人工耳蜗使用者中比较了aCAEP与eCAEP,两者P1-N1-P2记录清晰可靠且相关性良好。
  • 刺激水平降低时潜伏期延长而波幅下降。
  • eCAEP因直接刺激减少言语处理器干扰与延迟、记录与刺激设置更灵活,可能是验证皮层水平言语检测的更优选择,并可作为人工耳蜗调试的替代方法。

研究设计与样本

  • aCAEP用HEARLab系统在自由场以55 dB SPL呈现/m/、/g/、/t/、/s/四种言语Token记录。
  • eCAEP用Evoked Potential设备经MAX Programming Interface触发,以70 msec电脉冲串、0.9 Hz刺激耳蜗顶部(1)、中部(6)、底部(10或11)电极,在MCL及50% MCL下记录。

主要发现

  • aCAEP在言语Token /t/的100%受试者中记录到,/g/和/s/为95%,/m/为90%。
  • eCAEP在刺激顶部和中部电极时所有受试者均可识别N1和P2峰;同电极50% MCL时检出率降至85%。
  • P1峰在所有电极均较不明显。
  • 50% MCL较MCL整体潜伏期延长,相邻峰(P1-N1和N1-P2)波幅差异显著。
  • 四种言语Token的最大互相关均值在0.63至0.68范围;/g/的有效互相关数最多,/s/最少。

关键数值与推荐

  • aCAEP记录检出率:/t/为100%,/g/和/s/为95%,/m/为90%。
  • eCAEP在顶部和中部电极所有受试者均可识别N1和P2,50% MCL时检出率为85%。
  • 四种言语Token最大互相关均值范围为0.63至0.68。

临床怎么用

  • eCAEP可用于人工耳蜗调试中在皮层水平验证言语检测。
  • 刺激水平降低时需关注潜伏期延长与波幅下降的变化。
  • eCAEP因直接刺激干扰小、设置灵活,可作为aCAEP的替代记录方式。

局限性

  • P1峰在所有电极均较不明显,可能影响波形完整判读。
  • 50% MCL时eCAEP的N1和P2检出率降至85%。
  • 不同言语Token的有效互相关数量存在差异,/s/最少。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The primary goal of the study was to investigate electrical cortical auditory evoked potentials (eCAEPs) at maximum comfortable level (MCL) and 50% MCL on three cochlear implant (CI) electrodes and compare them with the acoustic CAEP (aCAEPs), in terms of the amplitude and latency of the P1-N1-P2 complex. This was achieved by comparing the eCAEP obtained with the method described and stimulating single electrodes, via the fitting software spanning the cochlear array and the aCAEP obtained using the HEARLab system at four speech tokens. DESIGN: Twenty MED-EL (MED-EL Medical Electronics, Innsbruck, Austria) CI adult users were tested. CAEP recording with HEARLab System was performed with speech tokens /m/, /g/, /t/, and /s/ in free field, presented at 55 dB SPL. eCAEPs were recorded with an Evoked Potential device triggered from the MAX Programming Interface (MED-EL Medical Devices) with 70 msec electrical burst at 0.9 Hz at the apical (1), middle (6), and basal (10 or 11) CI electrode at their MCL and 50% MCL. RESULTS: CAEP responses were recorded in 100% of the test subjects for the speech token /t/, 95% for the speech tokens /g/ and /s/, and 90% for the speech token /m/. For eCAEP recordings, in all subjects, it was possible to identify N1 and P2 peaks when stimulating the apical and middle electrodes. This incidence of detection decreased to an 85% chance of stimulation at 50% MCL on the same electrodes. A P1 peak was less evident for all electrodes. There was an overall increase in latency for stimulation at 50% MCL compared with MCL. There was a significant difference in the amplitude of adjacent peaks (P1-N1 and N1-P2) for 50% MCL compared with MCL. The mean of the maximum cross-correlation values were in the range of 0.63 to 0.68 for the four speech tokens. The distribution of the calculated time shift, where the maximum of the cross-correlation was found, was distributed between the speech tokens. The speech token /g/ had the highest number of valid cross-correlations, while the speech token /s/ had the lowest number. CONCLUSIONS: This study successfully compared aCAEP and eCAEP in CI users. Both acoustic and electrical P1-N1-P2 recordings obtained were clear and reliable, with good correlation. Latency increased with decreasing stimulation level, while amplitude decreased. eCAEP is potentially a better option to verify speech detection at the cortical level because it (1) uses direct stimulation and therefore creates less interference and delay of the sound processor and (2) creates more flexibility with the recording setup and stimulation setting. As such, eCAEP is an alternative method for CI optimization.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Távora-Vieira D, Mandruzzato G, Polak M, et al. Comparative Analysis of Cortical Auditory Evoked Potential in Cochlear Implant Users. Ear Hear. 2021. PMID: 34172688. doi:10.1097/AUD.0000000000001075

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。