耳蜗内阻抗测量检测人工耳蜗电极阵列移位
本研究基于100例HiFocus Mid-Scala人工耳蜗接受者的电场成像记录,计算电极阻抗和接入电阻,以CT扫描为金标准评估两者检测电极移位的可靠性。结果显示两者均可靠,但接入电阻的准确率显著更高,且能可靠检测移位位置,推荐作为检测电极阵列移位的方法。
要点
- 接入电阻检测电极移位的准确率显著高于电极阻抗。
- 接入电阻预测的移位位置与CT扫描结果显著相关。
- 电极阻抗与移位位置无显著相关,不推荐用于定位。
核心结论
- 电极阻抗和接入电阻均可可靠检测HiFocus Mid-Scala电极阵列的移位,但接入电阻准确率显著更优且能可靠检测移位位置。
- 推荐测量接入电阻作为检测电极阵列移位的方法。
研究设计与样本
- 研究纳入100例HiFocus Mid-Scala人工耳蜗接受者,利用电场成像记录计算电极阻抗和接入电阻。
- 以临床CT扫描为金标准,采用自助抽样法和受试者工作特征曲线分析分类移位与非移位。
主要发现
- 电极阻抗检测移位的准确率95%置信区间为77.8%至82.1%。
- 接入电阻检测移位的准确率95%置信区间为89.5%至91.2%。
- 接入电阻的最大偏差准确率显著大于电极阻抗的最大偏差准确率。
- 接入电阻预测的移位位置与CT扫描结果显著相关,而电极阻抗无显著相关。
关键数值与推荐
- 电极阻抗检测移位的准确率95%置信区间为77.8%至82.1%。
- 接入电阻检测移位的准确率95%置信区间为89.5%至91.2%。
- 接入电阻预测的移位位置与CT扫描结果显著相关。
- 电极阻抗与移位位置无显著相关。
临床怎么用
- 测量接入电阻可在电极插入后为外科医生提供即时反馈,有助于改进手术技巧并最终减少移位。
- 未来该测量可能实现插入过程中近实时监测电极阵列,帮助避免移位。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Misplacement of the electrode array is associated with impaired speech perception in patients with cochlear implants (CIs). Translocation of the electrode array is the most common misplacement. When a CI is translocated, it crosses the basilar membrane from the scala tympani into the scala vestibuli. The position of the implant can be determined on a postoperative CT scan. However, such a scan is not obtained routinely after CI insertion in many hospitals, due to radiation exposure and processing time. Previous studies have shown that impedance measures might provide information on the placement of the electrode arrays. The electrode impedance was measured by dividing the plateau voltage at the end of the first phase of the pulse by the injected current. The access resistance was calculated using the so-called access voltage at the first sampled time point after the start of the pulse divided by the injected current. In our study, we obtained the electrode impedance and the access resistance to detect electrode translocations using electrical field imaging. We have investigated how reliably these two measurements can detect electrode translocation, and which method performed best. DESIGN: We calculated the electrode impedances and access resistances using electrical field imaging recordings from 100 HiFocus Mid-Scala CI (Advanced Bionics, Sylmar, CA) recipients. We estimated the normal values of these two measurements as the baselines of the implant placed in the cochlea without translocation. Next, we calculated the maximal electrode impedance deviation and the maximal access-resistance deviation from the respective baselines as predictors of translocation. We classified these two predictors as translocations or nontranslocations based on the bootstrap sampling method and receiver operating characteristics curves analysis. The accuracy could be calculated by comparing those predictive results to a gold standard, namely the clinical CT scans. To determine which measurement more accurately detected translocation, the difference between the accuracies of the two measurements was calculated. RESULTS: Using the bootstrap sampling method and receiver operating characteristics-based optimized threshold criteria, the 95% confidence intervals of the accuracies of translocation detections ranged from 77.8% to 82.1% and from 89.5% to 91.2% for the electrode impedance and access resistance, respectively. The accuracies of the maximal access-resistance deviations were significantly larger than that of the maximal electrode impedance deviations. The location of the translocation as predicted by the access resistance was significantly correlated with the result derived from the CT scans. In contrast, no significant correlation was observed for the electrode impedance. CONCLUSIONS: Both the electrode impedance and access resistance proved reliable metrics to detect translocations for HiFocus Mid-Scala electrode arrays. The access resistance had, however, significantly better accuracy and it also reliably detected the electrode-location of translocations. The electrode impedance did not correlate significantly with the location of translocation. Measuring the access resistance is, therefore, the recommended method to detect electrode-array translocations. These measures can provide prompt feedback for surgeons after insertion, improving their surgical skills, and ultimately reducing the number of translocations. In the future, such measurements may allow near-real-time monitoring of the electrode array during insertion, helping to avoid translocations.
引用
- Dong Y, Briaire JJ, Siebrecht M, et al. Detection of Translocation of Cochlear Implant Electrode Arrays by Intracochlear Impedance Measurements. Ear Hear. 2021. PMID: 33974777. doi:10.1097/AUD.0000000000001033
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。