为人工耳蜗使用者个性化定制瞬态降噪算法设置

Personalizing Transient Noise Reduction Algorithm Settings for Cochlear Implant Users.
PubMed 文献 · Ear Hear · 2021-01-01

该研究在15名单侧人工耳蜗植入者中测试了宽带与多频带瞬态降噪算法参数个性化对 cafeteria 噪声下言语识别的影响。结果显示个性化多频带算法显著改善言语接收阈值1.3 dB,而个性化宽带算法显著恶化1.7 dB;主观评分及固定信噪比下词正确率均无显著变化。

要点

  • 个性化多频带瞬态降噪在言语接收阈值附近显著改善言语识别。
  • 个性化宽带瞬态降噪反而显著降低言语识别表现。
  • 在较优信噪比(SRT+3 dB)下各算法对词正确率无显著影响。

核心结论

  • 个性化多频带瞬态降噪设置仅在接近言语接收阈值的困难聆听条件下显著改善言语识别。
  • 在更有利的信噪比(SRT+3 dB)下该获益消失。
  • 组平均设置的多频带算法对言语识别无显著影响。

研究设计与样本

  • 研究纳入15名单侧人工耳蜗植入者,在食堂噪声(餐具碰撞声与多人交谈声组合)下测试。
  • 受试者可 individually 调整宽带与多频带瞬态降噪算法的衰减缩放因子与释放时间(τ)两个参数。
  • 以 Advanced Bionics 现行临床标准宽带瞬态降噪 SoundRelax(TNRbb-std)作为参照。
  • 采用自适应方法测定言语接收阈值(SRT)评估算法效果。
  • 在固定信噪比 SRT+3 dB 下进行主观评分(言语清晰度、自然度、聆听努力、噪声烦扰度)并测定词正确率。

主要发现

  • 个性化多频带瞬态降噪使 SRT 显著改善1.3 dB。
  • 个性化宽带瞬态降噪使 SRT 显著恶化1.7 dB。
  • 使用排除无获益受试者后确定的组平均设置,多频带算法对 SRT 不再有显著影响。
  • 标准宽带瞬态降噪 TNRbb-std 对言语识别无显著影响。
  • 所有算法对四项主观评分及 SRT+3 dB 下词正确率均无显著影响。

关键数值与推荐

  • 个性化多频带瞬态降噪:SRT 显著改善1.3 dB。
  • 个性化宽带瞬态降噪:SRT 显著恶化1.7 dB。
  • 组平均多频带设置:对 SRT 无显著影响。
  • TNRbb-std:对言语识别无显著影响。
  • SRT+3 dB 下词正确率:所有算法(包括个性化多频带)均无显著影响。

临床怎么用

  • 多频带瞬态降噪的参数个性化可能改善人工耳蜗使用者在困难噪声条件下的言语识别。
  • 宽带瞬态降噪的个性化参数设置可能损害言语识别,不宜随意个性化。
  • 现有 SoundRelax 为宽带滤波器且不支持参数个性化,未来可考虑升级为可个性化的多频带变体。

局限性

  • 研究仅纳入15名单侧人工耳蜗植入者,样本量有限。
  • 获益仅见于接近 SRT 的困难条件,在 SRT+3 dB 下未观察到获益。
  • 组平均设置未能复现个性化设置的获益。
  • 主观评分在所有测试条件下均无显著变化。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Speech understanding in noise is difficult for patients with a cochlear implant. One common and disruptive type of noise is transient noise. We have tested transient noise reduction (TNR) algorithms in cochlear implant users to investigate the merits of personalizing the noise reduction settings based on a subject's own preference. DESIGN: The effect of personalizing two parameters of a broadband and a multiband TNR algorithm (TNRbb and TNRmb, respectively) on speech recognition was tested in a group of 15 unilaterally implanted subjects in cafeteria noise. The noise consisted of a combination of clattering dishes and babble noise. Each participant could individually vary two parameters, namely the scaling factor of the attenuation and the release time (τ). The parameter τ represents the duration of the attenuation applied after a transient is detected. As a reference, the current clinical standard TNR "SoundRelax" from Advanced Bionics was tested (TNRbb-std). Effectiveness of the algorithms on speech recognition was evaluated adaptively by determining the speech reception threshold (SRT). Possible subjective benefits of the algorithms were assessed using a rating task at a fixed signal-to-noise ratio (SNR) of SRT + 3 dB. Rating was performed on four items, namely speech intelligibility, speech naturalness, listening effort, and annoyance of the noise. Word correct scores were determined at these fixed speech levels as well. RESULTS: The personalized TNRmb improved the SRT statistically significantly with 1.3 dB, while the personalized TNRbb degraded it significantly by 1.7 dB. For TNRmb, we attempted to further optimize its settings by determining a group-based setting, leaving out those subjects that did not experience a benefit from it. Using these group-based settings, however, TNRmb did not have a significant effect on the SRT any longer. TNRbb-std did not affect speech recognition significantly. No significant effects on subjective ratings were found for any of the items investigated. In addition, at a constant speech level of SRT + 3 dB, no effect of any of the algorithms was found on word correct scores, including TNRmb with personalized settings. CONCLUSIONS: Our study results indicate that personalizing noise reduction settings of a multiband TNR algorithm can significantly improve speech intelligibility in transient noise, but only under challenging listening conditions around the SRT. At more favorable SNRs (SRT + 3 dB), this benefit was lost. We hypothesize that TNRmb was beneficial at lower SNRs, because of more effective artifact detection under those conditions. Group-averaged settings of the multiband algorithm did not significantly affect speech recognition. TNRbb decreased speech recognition significantly using personalized parameter settings. Rating scores were not significantly affected by the algorithms under any condition tested. The currently available TNR algorithm for Advanced Bionics systems (SoundRelax) is a broadband filter that does not support personalization of its settings. Future iterations of this algorithm might benefit from upgrading it to a multiband variant with the option to personalize its parameter settings.