术中耳蜗电图与术后锥形束CT验证耳蜗微音电位振幅与植入电极位置的可重复性

Establishing Reproducibility and Correlation of Cochlear Microphonic Amplitude to Implant Electrode Position Using Intraoperative Electrocochleography and Postoperative Cone Beam Computed Tomography.
PubMed 文献 · Ear Hear · 2021-01-01

本研究纳入12例患者(14耳),在人工耳蜗电极插入过程中及插入后记录耳蜗微音电位(CM),并结合术后锥形束CT(CBCT)评估电极接触位置。结果显示术中插入角度与术后角度在两个振幅事件上显著相关(最大振幅:平均绝对误差29°,r=0.77,p=0.006;最大振幅10%:平均绝对误差52°,r=0.85,p=0.002),并据此构建了个体化耳蜗声学响应区域图。

要点

  • 术中与术后CM振幅对应的电极角度显著相关,支持记录可重复性
  • 最大振幅处平均绝对误差29°(r=0.77,p=0.006)
  • 最大振幅10%处平均绝对误差52°(r=0.85,p=0.002)

核心结论

  • 开发了一种新方法证明插入过程中从CI电极记录的CM反应具有可重复性。
  • 通过CM振幅与术后CBCT相关,可创建个体化CM反应图并将耳蜗分为声学响应与无响应区域。

研究设计与样本

  • 前瞻性研究,纳入12例患者(n=14耳),术中及插入后记录CM,术后行CBCT。
  • 术中采用0.5 kHz短音刺激从顶端电极接触记录;插入后从奇数编号电极接触(1-15)使用0.125至2 kHz频率记录。

主要发现

  • 术中插入与术后角度在最大振幅事件显著相关:平均绝对误差29°,r=0.77,p=0.006。
  • 术中插入与术后角度在最大振幅10%事件显著相关:平均绝对误差52°,r=0.85,p=0.002。

关键数值与推荐

  • 最大振幅:平均绝对误差29°,r=0.77,p=0.006。
  • 最大振幅10%:平均绝对误差52°,r=0.85,p=0.002。

临床怎么用

  • 术中CM记录结合术后CBCT可用于评估电极位置与声学响应区域。
  • 个体化耳蜗响应图可能有助于优化电极映射与放置,但需进一步验证其与响度分辨的关系。

局限性

  • 样本量较小(12例患者,14耳)。
  • 声学敏感区域电极接触与响度分辨改善的关联尚未验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The primary objective of this study was to establish the reproducibility of cochlear microphonic (CM) recordings obtained from a cochlear implant (CI) electrode contact during and immediately after insertion. This was achieved by evaluating the insertion angle and calculating the position of the apical electrode contact during insertion, using postoperative cone beam computed tomography (CBCT). The secondary objective was to create individualized patient maps of electrode contacts located within acoustically sensitive regions by correlating the CM amplitude to the electrode position determined using CBCT. METHODS: CMs were recorded from a CI electrode contact during and immediately after insertion in 12 patients (n = 14 ears). Intraoperative recordings were made for a 0.5 kHz tone burst stimulus and were recorded from the apical electrode contact. Postinsertion recordings were made from the odd-numbered electrode contacts (1-15) along the array, using a range of stimulus frequencies (from 0.125 to 2 kHz). The time point at which each electrode contact passed through the round window was noted throughout the insertion, and the CM amplitude at this point was correlated to postoperative CBCT. This correlation was then used to estimate the CM amplitude at particular points within the cochlea, which was in turn compared with the amplitudes recorded from each electrode postoperatively to assess the reproducibility of the recordings. RESULTS: Significant correlation was shown between intraoperative insertion and postinsertion angles at two amplitude events (maximum amplitude: 29° mean absolute error, r = 0.77, p = 0.006; 10% of maximum amplitude: 52° mean absolute error, r = 0.85, p = 0.002). CONCLUSION: We have developed a novel method to demonstrate the reproducibility of the CM responses recorded from a CI electrode during insertion. By correlating the CM amplitude with the postoperative CBCT, we have also been able to create individualized maps of CM responses, categorizing the cochlea into acoustically responsive and unresponsive regions. If the electrode contacts within the acoustically sensitive regions are shown to be associated with improved loudness discrimination, it could have implications for optimal electrode mapping and placement.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Soulby A, Connor S, Jiang D, et al. Establishing Reproducibility and Correlation of Cochlear Microphonic Amplitude to Implant Electrode Position Using Intraoperative Electrocochleography and Postoperative Cone Beam Computed Tomography. Ear Hear. 2021. PMID: 33813521. doi:10.1097/AUD.0000000000001010