文献速递咽喉与嗓音

喉部力学传感器在支撑喉镜下微喉手术中的前瞻性对照试验

Laryngeal Force Sensor for Suspension Microlaryngoscopy: A Prospective Controlled Trial.
PubMed 文献 · Otolaryngol Head Neck Surg · 2021-01-05

本研究在学术三级中心开展前瞻性对照试验,前100例手术中术者不知晓力学数据,后100例主动使用喉部力学传感器监测。结果显示主动监测组总冲量更低、喉外并发症更少,最大力可预测术后并发症,主动监测使并发症风险降低29.1%。

要点

  • 最大力是支撑喉镜术后并发症的独立预测因素。
  • 主动使用喉部力学传感器监测可降低术后并发症风险。
  • 主动监测还能减少异常EAT-10评分(吞咽障碍)的发生。

核心结论

  • 最大力可预测支撑喉镜微喉手术术后并发症的发生。
  • 主动使用喉部力学传感器监测能够减轻术中力并预防术后并发症。

研究设计与样本

  • 前瞻性对照试验,在学术三级中心开展。
  • 连续100例患者术者不知晓力学读数,随后100例术者主动使用喉部力学传感器监测系统。
  • 患者前瞻性入组并完成术前和术后问卷调查,评估舌痛、感觉异常、轻瘫、味觉障碍或吞咽困难的发生。

主要发现

  • 单因素分析中,主动监测组总冲量更低(P < .001),喉外并发症更少(P < .01)。
  • 多因素logistic回归中,最大力是术后并发症的显著预测变量(OR 1.08;95% CI 1.01-1.16;P = .02)。
  • 主动LFS监测使术后并发症发生可能性降低29.1%(95% CI 15.7%-42.4%;P < .001)。
  • 最大力(OR 1.11;95% CI 1.04-1.18;P < .01)和主动LFS监测(降低16.6%;95% CI 2.7%-30.5%;P = .02)可预测异常10项EAT评分。
  • 上述效应在术后第一次随访时仍持续:最大力(P = .04)和主动LFS监测(P = .01)对并发症;最大力(P = .01)和主动LFS监测(P = .01)对异常EAT-10评分。

临床怎么用

  • 在支撑喉镜微喉手术中可考虑使用喉部力学传感器实时监测力值。
  • 术者主动参考力学数据有助于减少总冲量和最大力,降低术后喉外并发症及吞咽障碍风险。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The laryngeal force sensor (LFS) provides real-time force data for suspension microlaryngoscopy. This study investigates whether active use of the LFS can prevent the development of complications. STUDY DESIGN: Prospective controlled trial. SETTING: Academic tertiary center. METHODS: The LFS and custom software were developed to track intraoperative force metrics. A consecutive series of 100 patients had force data collected with operating surgeons blinded to intraoperative readings. The subsequent 100 patients had surgeons actively use the LFS monitoring system. Patients were prospectively enrolled, completing pre- and postoperative surveys to assess the development of tongue pain, paresthesia, paresis, dysgeusia, or dysphagia. RESULTS: On univariate analysis, the active monitoring group had lower total impulse (P < .001) and fewer extralaryngeal complications (P < .01). On multiple logistic regression, maximum force (odds ratio [OR], 1.08; 95% CI, 1.01-1.16; P = .02) was a significant predictive variable for the development of postoperative complications. Similarly, active LFS monitoring showed a 29.1% (95% CI, 15.7%-42.4%; P < .001) decrease in the likelihood of developing postoperative complications. These effects persisted at the first postoperative visit for maximum force (P = .04) and active LFS monitoring (P = .01). Maximum force (OR, 1.11; 95% CI, 1.04-1.18; P < .01) and active LFS monitoring (16.6%; 95% CI, 2.7%-30.5%; P = .02) were also predictive for the development of an abnormal 10-item Eating Assessment Tool score. These effects also persisted at the first postoperative visit for maximum force (P = .01) and active LFS monitoring (P = .01). CONCLUSION: Maximum force is predictive of the development postoperative complications. Active monitoring with the LFS is able to mitigate these forces and prevent postoperative complications. LEVEL OF EVIDENCE: 2.

来源:Otolaryngol Head Neck Surg,查看原文。版权归原作者所有。

引用

  1. Feng AL, Puka E, Ciaramella A, et al. Laryngeal Force Sensor for Suspension Microlaryngoscopy: A Prospective Controlled Trial. Otolaryngol Head Neck Surg. 2021. PMID: 33399514. doi:10.1177/0194599820982635

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。