影响成人人工耳蜗植入者言语感知的因素
横断面研究纳入129例经验丰富的成人人工耳蜗植入者,分析传记、听力及电极位置因素对言语感知的影响。结果显示,在矫正相关因素后,预弯电极较直电极在CVC50音素感知上高11.8个百分点(95%CI:1.4-20.4%;p=0.03)。
要点
- 预弯电极较直电极显著提高CVC50音素感知得分。
- 直电极组中植入年龄较低与CVC50得分较高相关。
- 直电极保留残余听力效果差,术前中位88 dB,术后122 dB。
核心结论
- 预弯电极植入带来显著更高的言语感知结果,独立于传记因素、听力因素和阶区位置。
研究设计与样本
- 横断面研究,纳入129例语后聋经验丰富的成人人工耳蜗植入者,均使用同一制造商设备。
- 参与者接受预弯电极(N=85)或直电极(N=44),数据收集于2016年12月至2018年1月。
- 言语感知测量采用CVC50、CVC65和数字噪声测试,电极位置通过超高分辨率CT扫描评估。
主要发现
- 直电极组中,较低的植入年龄与较高的CVC50音素得分相关。
- 其他传记变量与言语感知不相关。
- 预弯电极且术前听力阈值较差者,在所有言语感知结果上优于直电极且术前听力阈值相对较好者。
- 矫正后,预弯电极较直电极CVC50音素感知得分高11.8个百分点(95%CI:1.4-20.4%;p=0.03)。
- 直电极保留残余听力效果差,术前中位残余听力阈值88 dB,术后122 dB。
关键数值与推荐
- 预弯电极较直电极CVC50音素感知得分高11.8个百分点(95%CI:1.4-20.4%;p=0.03)。
- 直电极术前中位残余听力阈值88 dB,术后122 dB。
临床怎么用
- 成人人工耳蜗植入选择预弯电极可能获得更好的言语感知结果。
- 直电极对残余听力的保留效果有限,术前咨询需考虑此因素。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: The primary objective of this study is to identify the biographic, audiologic, and electrode position factors that influence speech perception performance in adult cochlear implant (CI) recipients implanted with a device from a single manufacturer. The secondary objective is to investigate the independent association of the type of electrode (precurved or straight) with speech perception. DESIGN: In a cross-sectional study design, speech perception measures and ultrahigh-resolution computed tomography scans were performed in 129 experienced CI recipients with a postlingual onset of hearing loss. Data were collected between December 2016 and January 2018 in the Radboud University Medical Center, Nijmegen, the Netherlands. The participants received either a precurved electrode (N = 85) or a straight electrode (N = 44), all from the same manufacturer. The biographic variables evaluated were age at implantation, level of education, and years of hearing loss. The audiometric factors explored were preoperative and postoperative pure-tone average residual hearing and preoperative speech perception score. The electrode position factors analyzed, as measured from images obtained with the ultrahigh-resolution computed tomography scan, were the scalar location, angular insertion depth of the basal and apical electrode contacts, and the wrapping factor (i.e., electrode-to-modiolus distance), as well as the type of electrode used. These 11 variables were tested for their effect on three speech perception outcomes: consonant-vowel-consonant words in quiet tests at 50 dB SPL (CVC50) and 65 dB SPL (CVC65), and the digits-in-noise test. RESULTS: A lower age at implantation was correlated with a higher CVC50 phoneme score in the straight electrode group. Other biographic variables did not correlate with speech perception. Furthermore, participants implanted with a precurved electrode and who had poor preoperative hearing thresholds performed better in all speech perception outcomes than the participants implanted with a straight electrode and relatively better preoperative hearing thresholds. After correcting for biographic factors, audiometric variables, and scalar location, we showed that the precurved electrode led to an 11.8 percentage points (95% confidence interval: 1.4-20.4%; p = 0.03) higher perception score for the CVC50 phonemes compared with the straight electrode. Furthermore, contrary to our initial expectations, the preservation of residual hearing with the straight electrode was poor, as the median preoperative and the postoperative residual hearing thresholds for the straight electrode were 88 and 122 dB, respectively. CONCLUSIONS: Cochlear implantation with a precurved electrode results in a significantly higher speech perception outcome, independent of biographic factors, audiometric factors, and scalar location.
引用
- Heutink F, Verbist BM, van der Woude WJ, et al. Factors Influencing Speech Perception in Adults With a Cochlear Implant. Ear Hear. 2021. PMID: 33480623. doi:10.1097/AUD.0000000000000988
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。