侧壁与蜗轴周电极阵列的阶间移位比较:一项Meta分析
该系统综述纳入33项原始研究,其中5项队列研究比较了侧壁与蜗轴周电极阵列的阶间移位率。Meta分析显示侧壁阵列移位率显著更低(7% vs. 43%,合并OR=0.12),移位与言语感知评分负相关(加权均值41% vs. 55%),蜗轴周阵列尖端折叠更常见(X2=6.8,P<.01)。
要点
- 侧壁电极阵列阶间移位率显著低于蜗轴周阵列(7% vs. 43%)。
- 阶间移位与较低言语感知评分相关(41% vs. 55%)。
- 蜗轴周阵列尖端折叠发生率更高(X2=6.8,P<.01)。
核心结论
- 蜗轴周电极阵列的阶间移位和尖端折叠发生率高于侧壁阵列。
- 阶间移位对人工耳蜗植入后的听力有负面影响。
- 若目标是减少临床相关的耳蜗内创伤,侧壁阵列是首选。
研究设计与样本
- 系统综述,检索PubMed、Embase和Cochrane数据库近11年文献,未设其他限制。
- 纳入33项原始研究,均为非随机对照试验。
- Meta分析纳入5项比较侧壁与蜗轴周阵列阶间移位的队列研究。
主要发现
- 侧壁阵列阶间移位率为7%,蜗轴周阵列为43%,合并OR=0.12。
- 阶间移位与言语感知评分负相关,加权均值分别为41%和55%。
- 蜗轴周阵列尖端折叠更频繁(X2=6.8,P<.01)。
临床怎么用
- 为减少耳蜗内创伤,人工耳蜗植入可优先选择侧壁电极阵列。
- 术后CT评估阶间移位和尖端折叠有助于了解电极位置。
局限性
- 纳入研究均为非随机对照试验。
- Meta分析仅基于5项队列研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES/HYPOTHESIS: Two types of electrode arrays for cochlear implants (CIs) are distinguished: lateral wall and perimodiolar. Scalar translocation of the array can lead to intracochlear trauma by penetrating from the scala tympani into the scala vestibuli or scala media, potentially negatively affecting hearing performance of CI users. This systematic review compares the lateral wall and perimodiolar arrays with respect to scalar translocation. STUDY DESIGN: Systematic review. METHODS: PubMed, Embase, and Cochrane databases were reviewed for studies published within the last 11 years. No other limitations were set. All studies with original data that evaluated the occurrence of scalar translocation or tip fold-over (TF) with postoperative computed tomography (CT) following primary cochlear implantation in bilateral sensorineuronal hearing loss patients were considered to be eligible. Data were extracted independently by two reviewers. RESULTS: We included 33 studies, of which none were randomized controlled trials. Meta-analysis of five cohort studies comparing scalar translocation between lateral wall and perimodiolar arrays showed that lateral wall arrays have significantly lower translocation rates (7% vs. 43%; pooled odds ratio = 0.12). Translocation was negatively associated with speech perception scores (weighted mean 41% vs. 55%). Tip fold-over of the array was more frequent with perimodiolar arrays (X2 = 6.8, P < .01). CONCLUSIONS: Scalar translocation and tip fold-overs occurred more frequently with perimodiolar arrays than with lateral wall arrays. In addition, translocation of the array negatively affects hearing with the cochlear implant. Therefore, if one aims to minimize clinically relevant intracochlear trauma, lateral wall arrays would be the preferred option for cochlear implantation. Laryngoscope, 131:1358-1368, 2021.
引用
- Jwair S, Prins A, Wegner I, et al. Scalar Translocation Comparison Between Lateral Wall and Perimodiolar Cochlear Implant Arrays - A Meta-Analysis. Laryngoscope. 2020. PMID: 33159469. doi:10.1002/lary.29224