文献速递耳科与耳神经

振动声桥圆窗耦合的长期稳定性与安全性

Long-Term Stability and Safety of the Soundbridge Coupled to the Round Window.
PubMed 文献 · Laryngoscope · 2020-11-19

该回顾性研究评估了2013至2019年间46例接受振动声桥(VSB)圆窗(RW)耦合植入患者的听力学结果与安全性,长期随访平均31.43个月(13-75个月)。结果显示无胆脂瘤患者无需修正手术,有胆脂瘤患者修正率为15.2%,残余听力未受影响,Freiburger单音节词识别在65 dB和80 dB均显著改善(P < .0001)且长期稳定。

要点

  • VSB圆窗耦合可安全用于传导性或混合性听力损失患者。
  • 有胆脂瘤(伴根治腔)者修正手术风险较高,需额外告知并发症。
  • 术后言语识别改善显著且长期稳定,残余听力未受影响。

核心结论

  • VSB圆窗耦合植入对传导性或混合性听力损失患者安全,听力改善长期稳定。
  • 修正率与胆脂瘤(伴根治腔)基础病变直接相关。

研究设计与样本

  • 回顾性研究,纳入2013至2019年奥地利St. Pölten大学医院46例VSB圆窗耦合患者。
  • 样本分为随访小于12个月组和大于12个月组,并按有无胆脂瘤分别分析累积生存结局。

主要发现

  • 46例患者平均长期随访31.43个月(13-75个月)。
  • 有胆脂瘤患者中7例(15.2%)需修正手术,无胆脂瘤患者无修正手术(0%)。
  • 残余听力未受VSB手术影响。
  • Freiburger单音节词识别在65 dB(P < .0001)和80 dB(P < .0001)显著改善,且长期随访稳定。

关键数值与推荐

  • 听力改善在长达74个月的长期随访中保持稳定。
  • 有胆脂瘤(伴根治腔)患者需额外告知潜在并发症。

临床怎么用

  • 对于传导性或混合性听力损失,可考虑VSB圆窗耦合植入。
  • 对合并胆脂瘤(伴根治腔)者应加强并发症告知与随访。

局限性

  • 证据等级为4(病例系列)。
  • 回顾性设计,样本量46例。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The objective of the study was to demonstrate the long-term outcomes of patients implanted with the active middle ear implant (AMEI) Vibrant Soundbridge (VSB) through coupling the floating mass transducer (FMT) to the round window (RW). METHODS: This retrospective study evaluated the short- and long-term clinical performance (audiological outcomes) and safety (revisions/explantations) of the VSB coupled to the RW between 2013 and 2019 at the St. Pölten University Hospital, Austria. For the outcome analysis, the sample was divided into a short-term examination group followed up for less than 12 months (<12 months) and a long-term examination group followed up for more than 12 months (>12 months). Cumulative survival outcomes were separately analyzed for subjects with and without cholesteatoma. RESULTS: 46 patients with an average long-term follow-up period of 31.43 months (13-75 months) were investigated. Complications requiring revision surgery were reported in total in seven patients with cholesteatoma (15.2%) and none in subjects without cholesteatoma (0%). Residual hearing was not affected by VSB surgery. Word understanding on the Freiburger monosyllabic speech test improved significantly at 65 dB (P < .0001) and 80 dB (P < .0001), and these outcomes were stable for long-term follow up. CONCLUSION: The VSB coupled to the RW is a safe implantation method for patients with conductive or mixed hearing loss. Hearing improvement was stable for the long-term follow-up up to 74 months. The revision rates are directly related to the underlying pathology of cholesteatoma (with radical cavity); thus, this special cohort requires additional counseling on potential complications. LEVEL OF EVIDENCE: 4 (Case-series) Laryngoscope, 131:E1434-E1442, 2021.

来源:Laryngoscope,查看原文。版权归原作者所有。

引用

  1. Sprinzl GM, Schoerg P, Muck S, et al. Long-Term Stability and Safety of the Soundbridge Coupled to the Round Window. Laryngoscope. 2020. PMID: 33210744. doi:10.1002/lary.29269

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。