其中患者不良事件40起(42.1%),操作者不良事件2起(2.1%),设备故障53起(55.8%)。患者不良事件以激光光纤脱落(23起,57.5%)、烧伤(8起,20%)和瘢痕(5起,12.5%)最常见;设备故障以光纤断裂(26起,49.1%)和 flare(12起,22.6%)最常见。作者呼吁标准化报告和多中心研究以准确估计发生率。

要点

  • 上气道激光手术不良事件中设备故障占比最高(55.8%),其次为患者不良事件(42.1%)。
  • 患者最常见不良事件为激光光纤脱落(57.5%)、烧伤(20%)和瘢痕(12.5%)。
  • 设备故障以光纤断裂(49.1%)和 flare(22.6%)为主,操作者不良事件仅2起眼暴露。

核心结论

  • 上气道激光手术与患者不良事件、操作者暴露及设备故障相关,当前报告存在不足。

研究设计与样本

  • 查询美国FDA MAUDE数据库2010年1月至2020年3月手术激光不良事件报告,提取上气道相关数据。

主要发现

  • 95起事件中,患者不良事件40起(42.1%),操作者不良事件2起(2.1%),设备故障53起(55.8%)。
  • 患者不良事件最常见为激光光纤脱落23起(57.5%)、烧伤8起(20%)、瘢痕5起(12.5%)。
  • 操作者不良事件为2起经未过滤显微镜镜片的眼暴露。
  • 设备故障最常见为光纤断裂26起(49.1%)和 flare 12起(22.6%)。

临床怎么用

  • 临床应关注上气道激光手术中光纤脱落、烧伤和瘢痕等患者风险及光纤断裂等设备故障。
  • 需标准化报告和多中心研究以更好理解不良事件并准确估计发生率。

局限性

  • 原文指出当前报告存在不足,需标准化报告和多中心研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Surgical lasers are used extensively in head and neck surgery. Laser use in the upper airway offers many advantages but also presents risks to patients and operators that are not reported comprehensively. This study aims to summarize device malfunctions, patient complications, and subsequent interventions related to laser use in the upper airway. METHODS: The US Food and Drug Administration's Manufacturer and User Facility Device Experience database was queried for reports of surgical laser adverse events from January 2010 to March 2020. Data were extracted from reports pertaining to the upper airway. RESULTS: Sixty-two reports involving upper airway laser use in an operating room were identified, from which 95 events were extracted. Of these, 40 (42.1%) were adverse events to patients, 2 (2.1%) adverse events to operators, and 53 (55.8%) device malfunctions. Dislodgement of laser fiber in the airway (23 [57.5%]), burn (8 [20%]), and scar (5 [12.5%]) were the most common adverse events to patients. Two incidents of eye exposure through unfiltered microscope lenses were the only adverse events to operators. Fiber break (26 [49.1%]) and flare (12 [22.6%]) were the most common device malfunctions. DISCUSSION: Surgical lasers have demonstrated utility in head and neck surgery but are associated with risks. This study discusses adverse events and device malfunctions associated with airway laser surgery and emphasizes shortcomings in current reporting. IMPLICATIONS FOR PRACTICE: Standardized reporting and multi-institutional research are needed to better understand adverse events related to surgical laser use and to allow accurate estimation of their prevalence.