仿耳廓麦克风指向性改善双耳人工耳蜗使用者的声源定位与辨别
结果显示PI模式显著改善声源定位与辨别,前后混淆率从47%降至35%(p=0.03),90°与270°方位的最小可听角从75.5°降至37.7°(p<0.001),动态跟踪呈非显著改善趋势。
要点
- PI模式使前后混淆率从47%降至35%(p=0.03)。
- 仅PI模式能辨别正侧方(90°与270°)声源。
- 90°与270°最小可听角从75.5°降至37.7°(p<0.001)。
核心结论
- PI模式可改善双耳人工耳蜗使用者的空间听觉表现,主要源于前后方向辨别能力的提升。
研究设计与样本
- 每名受试者在水平面完成静态与动态声场空间听觉测试,静态测试包括绝对声源定位测试与八个方位的最小可听角测试。
- 动态声源跟踪能力通过受试者正确指示沿两条圆形路径移动的刺激方向来评估。
主要发现
- PI模式带来统计学显著的声源定位与辨别改善,最大获益为前后混淆减少。
- 前后混淆率从OMNI模式的47%降至PI模式的35%(p=0.03)。
- 正侧方(90°与270°方位)声源辨别仅在PI模式下可行。
- 90°与270°方位的平均最小可听角从75.5°降至37.7°(p<0.001)。
- 两条测试轨迹的动态声源跟踪能力均呈非显著改善趋势(p=0.34与p=0.27)。
临床怎么用
- 对双耳人工耳蜗使用者可考虑启用仿耳廓麦克风指向性模式以改善前后方向辨别与侧方定位。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: To compare the sound-source localization, discrimination, and tracking performance of bilateral cochlear implant users with omnidirectional (OMNI) and pinna-imitating (PI) microphone directionality modes. DESIGN: Twelve experienced bilateral cochlear implant users participated in the study. Their audio processors were fitted with two different programs featuring either the OMNI or PI mode. Each subject performed static and dynamic sound field spatial hearing tests in the horizontal plane. The static tests consisted of an absolute sound localization test and a minimum audible angle test, which was measured at eight azimuth directions. Dynamic sound tracking ability was evaluated by the subject correctly indicating the direction of a moving stimulus along two circular paths around the subject. RESULTS: PI mode led to statistically significant sound localization and discrimination improvements. For static sound localization, the greatest benefit was a reduction in the number of front-back confusions. The front-back confusion rate was reduced from 47% with OMNI mode to 35% with PI mode (p = 0.03). The ability to discriminate sound sources straight to the sides (90° and 270° angle) was only possible with PI mode. The averaged minimum audible angle value for the 90° and 270° angle positions decreased from a 75.5° to a 37.7° angle when PI mode was used (p < 0.001). Furthermore, a non-significant trend towards an improvement in the ability to track moving sound sources was observed for both trajectories tested (p = 0.34 and p = 0.27). CONCLUSIONS: Our results demonstrate that PI mode can lead to improved spatial hearing performance in bilateral cochlear implant users, mainly as a consequence of improved front-back discrimination with PI mode.
引用
- Fischer T, Schmid C, Kompis M, et al. Pinna-Imitating Microphone Directionality Improves Sound Localization and Discrimination in Bilateral Cochlear Implant Users. Ear Hear. 2021. PMID: 32701730. doi:10.1097/AUD.0000000000000912
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。