上气道刺激的气道反应前瞻性测定:高级设备滴定的新机会
结果显示标准设置(+-+)在43.8%患者中引起最大RP面积变化,而超过半数患者(56.2%)对替代设置反应更大,RP与RG扩张无相关性。
要点
- 标准设置(+-+)在43.8%患者中引起最大RP面积变化。
- 56.2%患者对替代电极设置反应更大,提示个体化滴定可能必要。
- RP与RG扩张无相关性,需分别评估不同解剖区域反应。
核心结论
- RP扩张程度在患者和电极设置间存在差异,标准设置仅在部分患者中引起最大RP变化。
主要发现
- 标准设置(+-+)在43.8%患者中引起最大RP面积变化,替代设置(---和o-o各18.8%,-o-12.5%,-+-6.3%)在其余患者中引起最大变化。
- 多数患者RP面积变化为20%-70%(43.8%)或无扩大(<20%,43.8%),少数(12.5%)有显著扩大(>70%)。
- RP与RG扩张无相关性。
关键数值与推荐
- 功能阈值下记录5种电极配置的RP和RG区域横截面积变化百分比,由两名盲法评估者测量。
- 根据5种设置的平均反应分类腭舌耦合程度:无扩大(<20%)、部分扩大(20%-70%)、显著扩大(>70%)。
临床怎么用
- UAS设备滴定时可考虑评估多种电极配置下的气道反应,而非仅使用标准设置。
- 未来需研究基于RP扩张选择设置是否能改善结局。
局限性
- 未报告长期临床结局与设置选择的关系。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES/HYPOTHESIS: Response to upper airway stimulation (UAS) is associated with the degree of airway opening during stimulation. UAS programming may affect this opening. The objective of this study was to examine airway changes in response to five different electrode configurations programmable within the Inspire UAS system. STUDY DESIGN: Prospective single-arm cohort study. METHODS: Subjects who underwent UAS implantation were recruited for a prospective single-arm cohort study during UAS device activation. Functional thresholds were recorded for all settings. Awake nasopharyngoscopy was performed to examine the retropalatal (RP) and retroglossal (RG) regions at rest and during activation with all settings at their functional thresholds. Cross-sectional measurements were made by two blinded reviewers and reported as percent change in airway size. RESULTS: Sixteen patients were included. The standard setting (+-+) resulted in the greatest change in RP area in 43.8% of patients. An alternative setting resulted in greatest change in 56.2% of patients (--- and o-o in 18.8% each, -o- in 12.5%, and -+- in 6.3% of patients). Average response to all five settings was utilized to classify degree of palatoglossal coupling. Most patients had some enlargement (20%-70% change in RP area, 43.8%) or no enlargement (<20% change, 43.8%), whereas a minority of patients (12.5%) had marked enlargement (>70% change). RP and RG expansion were not correlated. CONCLUSION: Degree of RP expansion varied among patients and settings. Although the standard setting resulted in greatest RP change in a plurality of patients, over half had a greater response to an alternative setting. Future studies should address whether choice of setting based on RP expansion results in improved outcomes. LEVEL OF EVIDENCE: 4 Laryngoscope, 131:218-223, 2021.
引用
- Jenks CM, Yu JL, Schmitt KA, et al. Prospective Determination of Airway Response to Upper Airway Stimulation: A New Opportunity for Advanced Device Titration. Laryngoscope. 2020. PMID: 32557705. doi:10.1002/lary.28758
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。