文献速递耳科与耳神经

diametric 磁铁人工耳蜗对磁共振成像可及性、安全性和临床诊疗的影响

Impact of Cochlear Implant With Diametric Magnet on Imaging Access, Safety, and Clinical Care.
PubMed 文献 · Laryngoscope · 2020-06-22

回顾性分析7例植入 diametric 磁铁人工耳蜗的患儿,共进行17次1.5T或3.0T磁共振检查(无头部固定)。76%检查在清醒状态下完成,无设备相关不适或磁铁相关并发症,93%脑部检查具临床价值。结论认为该设计提高了磁共振可及性并减少镇静需求。

要点

  • diametric 磁铁人工耳蜗可在不取磁铁下行1.5T或3.0T磁共振。
  • 多数检查无需镇静,无磁铁移位或不适相关并发症。
  • 伪影存在时多数脑部磁共振仍具临床价值。

核心结论

  • diametric 磁铁设计使磁铁在位时可行磁共振检查,避免了可移除轴向磁铁相关的疼痛和移位风险。
  • 该设计减少了磁共振相关并发症,因扭矩方向不垂直于磁铁平面且磁铁被牢固固定。
  • 该装置提高了磁共振可及性并降低了对镇静或麻醉的需求。

研究设计与样本

  • 回顾性病历审查,附加三级儿童医院磁共振资料回顾。
  • 7例患者,平均年龄8.4岁(范围1.3-19岁),磁铁在位期间接受磁共振。
  • 干预为一次或多次1.5T或3.0T磁共振检查,不使用头部包裹。

主要发现

  • 7例人工耳蜗使用者共接受17次1.5T或3.0T磁共振检查(磁铁在位)。
  • 17次磁共振中13次(76%)在清醒患者中完成。
  • 无患者出现设备相关不适,无磁铁相关并发症发生。
  • 14次脑部检查中13次(93%)尽管有伪影仍具临床价值。

临床怎么用

  • 对于植入 diametric 磁铁人工耳蜗的患者,可考虑在磁铁在位下行1.5T或3.0T磁共振检查而无需头部包裹。
  • 该装置可能减少磁共振检查中对镇静或麻醉的需求。

局限性

  • 回顾性设计,证据等级为4级。
  • 样本量小(7例患者,17次磁共振)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES/HYPOTHESIS: Review safety and effectiveness of magnetic resonance imaging (MRI) of patients implanted with a cochlear implant (CI) containing a diametric magnet housed within the undersurface of the device. STUDY DESIGN: Retrospective chart review with additional review of MRI at a tertiary-care children's hospital. METHODS: Seven patients with mean age of 8.4 years (range = 1.3-19 years) with a diametric magnet in situ during MRI. The intervention comprised one or more sessions of 1.5 T or 3.0 T MRI without a head wrap. The main outcome measures were the occurrence of magnet-related complications including discomfort and magnet displacement, use of anesthesia or sedation, and clinical usefulness of MRI studies. RESULTS: Seven CI recipients underwent 17 episodes of 1.5 or 3.0 T MRI with an in situ diametric magnet. Thirteen of 17 (76%) MRI sessions were completed in awake patients. No patients had device-related discomfort. No magnet-related complications occurred. Thirteen of 14 (93%) brain studies were clinically useful despite artifacts. CONCLUSIONS: The diametric magnet enabled MRI with magnet in situ without the discomfort or magnet displacement associated with removable axial magnets. The reduction in MRI magnet-related complications occurred because torque is not directed perpendicular and outward from the plane of the magnet, and the magnet is securely contained within its housing. The design of this device increased access and reduced the need for sedation or anesthesia. LEVEL OF EVIDENCE: 4 Laryngoscope, 131:E952-E956, 2021.

来源:Laryngoscope,查看原文。版权归原作者所有。

引用

  1. Young NM, Hoff SR, Ryan M, et al. Impact of Cochlear Implant With Diametric Magnet on Imaging Access, Safety, and Clinical Care. Laryngoscope. 2020. PMID: 32569426. doi:10.1002/lary.28854

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。