儿童人工耳蜗植入深度测量:耳蜗视图X线摄影的可靠性
本研究评估了耳蜗视图X线摄影测量婴幼儿及儿童人工耳蜗植入深度的评分者间与评分者内可靠性。两位耳蜗植入外科医生和两位放射科医生独立测量角度植入深度,6周后重复评估。评分者间与评分者内ICC在0.81至0.96之间,表明可靠性极好。
要点
- 耳蜗视图X线摄影测量儿童人工耳蜗植入深度具有极好的评分者间和评分者内可靠性。
- 49耳(86%)成功获得耳蜗视图X线片,中位角度植入深度为381°。
- 6周后重复测量中位深度为382°,ICC为0.81–0.96。
核心结论
- 术后耳蜗视图X线摄影是测量婴幼儿及儿童人工耳蜗植入深度的可靠工具。
研究设计与样本
- 前瞻性观察性研究,纳入11个月至20岁(中位4岁,IQR 1-8岁)接受人工耳蜗植入的儿童,研究期1年。
- 术后第1天行耳蜗视图成像,每耳选1张去标识化图像,由2名耳蜗植入外科医生和2名放射科医生独立评估角度植入深度,6周后重复评估。
主要发现
- 第一轮测量中位角度植入深度为381°(最小272°,最大450°,IQR 360°-395°)。
- 6周后重复测量中位深度为382°(最小272°,最大449°,IQR 360°-397°)。
- 评分者间和评分者内可靠性ICC在0.81至0.96之间,表明可靠性极好。
临床怎么用
- 术后耳蜗视图X线摄影可用于可靠测量儿童人工耳蜗植入深度。
- 需进一步研究确定术中获取的耳蜗视图X线片的可靠性。
局限性
- 需进一步研究确定术中获取的耳蜗视图X线片在该人群中的可靠性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Cochlear implant depth of insertion affects audiologic outcomes and can be measured in adults using plain films obtained in the "cochlear view." The objective of this study was to assess interrater and intrarater reliability of measuring depth of insertion using cochlear view radiography. STUDY DESIGN: Prospective, observational. SETTING: Tertiary referral pediatric hospital. SUBJECTS AND METHODS: Patients aged 11 months to 20 years (median, 4 years; interquartile range [IQR], 1-8 years) undergoing cochlear implantation at our institution were studied over 1 year. Children underwent cochlear view imaging on postoperative day 1. Films were deidentified and 1 image per ear was selected. Two cochlear implant surgeons and 2 radiologists evaluated each image and determined angular depth of insertion. Images were re-reviewed 6 weeks later by all raters. Inter- and intrarater reliability were calculated with intraclass correlation coefficients (ICCs). RESULTS: Fifty-seven ears were imaged from 42 children. Forty-nine ears (86%) had successful cochlear view x-rays. Median angular depth of insertion was 381° (minimum, 272°; maximum, 450°; IQR, 360°-395°) during the first round of measurement. Measurements of the same images reviewed 6 weeks later showed median depth of insertion of 382° (minimum, 272°; maximum, 449°; IQR, 360°-397°). Interrater and intrarater reliability ICCs ranged between 0.81 and 0.96, indicating excellent reliability. CONCLUSIONS: Postoperative cochlear view radiography is a reliable tool for measurement of cochlear implant depth of insertion in infants and children. Further studies are needed to determine reliability of intraoperatively obtained cochlear view radiographs in this population.
引用
- Noble AR, Christianson E, Norton SJ, et al. Reliability of Measuring Insertion Depth in Cochlear Implanted Infants and Children Using Cochlear View Radiography. Otolaryngol Head Neck Surg. 2020. PMID: 32450736. doi:10.1177/0194599820921857
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。