电凝功率设置对扁桃体切除术手术烟雾颗粒浓度的影响

Effect of Electrocautery Settings on Particulate Concentrations in Surgical Plume during Tonsillectomy.
PubMed 文献 · Otolaryngol Head Neck Surg · 2020-03-31

该横断面研究在36例儿童扁桃体电凝切除术中,用外科医生佩戴的便携式颗粒计数仪采样,比较12 W与20 W电凝、有无排烟系统时呼吸区颗粒浓度。结果显示20 W无排烟时颗粒浓度约为12 W的9.5倍以上,使用排烟系统后降至3.3倍。

要点

  • 20 W较12 W无排烟时颗粒浓度高逾9.5倍,排烟系统可将其降至3.3倍
  • 低功率电凝联合排烟系统可显著降低手术烟雾暴露

核心结论

  • 儿童扁桃体电凝切除术中,20 W较12 W无排烟时空气中颗粒浓度高逾9.5倍,使用排烟系统后降至3.3倍。
  • 低功率电凝配合排烟系统可显著降低患者、术者及手术室人员的手术烟雾暴露。

研究设计与样本

  • 横断面研究,在三级医疗中心开展。
  • 共纳入36例儿童,均以单极电凝完成全扁桃体切除术,用术者佩戴的便携式颗粒计数仪采样空气。
  • 分为12 W加排烟系统、20 W加排烟系统、12 W无排烟、20 W无排烟四组,每组9例。

主要发现

  • 呼吸区平均颗粒数浓度:12 W加排烟系统为1661 particles/cm3。
  • 20 W加排烟系统为5515 particles/cm3,12 W无排烟为8208 particles/cm3。
  • 20 W无排烟为78,506 particles/cm3,四组间颗粒数浓度差异有统计学意义。
  • 颗粒数浓度与电凝时间的相关性在12 W加排烟组为中等,其余组为负相关。

临床怎么用

  • 儿童扁桃体电凝切除术中建议采用较低电凝功率设置并配合排烟系统以降低手术烟雾暴露。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: To describe the effect of monopolar electrocautery (EC) settings on surgical plume particulate concentration during pediatric tonsillectomy. STUDY DESIGN: Cross-sectional study. SETTING: Tertiary medical center. SUBJECTS AND METHODS: During total tonsillectomy exclusively performed with EC, air was sampled with a surgeon-worn portable particle counter. The airborne mean and maximum particle concentrations were compared for tonsillectomy performed with EC at 12 W vs 20 W, with smoke evacuation system (SES) and no smoke evacuation (NS). RESULTS: A total of 36 children were included in this analysis: 9 cases with EC at 12 W and SES (12SES), 9 cases with EC at 20 W and SES (20SES), 9 cases with EC at 12 W without SES (12NS), and 9 cases with EC at 20 W without SES (20NS). Mean particle number concentration in the breathing zone during tonsillectomy was 1661 particles/cm3 for 12SES, 5515 particles/cm3 for 20SES, 8208 particles/cm3 for 12NS, and 78,506 particles/cm3 for 20NS. There was a statistically significant difference in the particle number concentrations among the 4 groups. The correlation between the particle number concentration and EC time was either moderate (for 12SES) or negative (for remaining groups). CONCLUSION: Airborne particle concentrations during tonsillectomy are over 9.5 times higher when EC is set at 20 W vs 12 W with NS, which is mitigated to 3.3 times with SES. Applying lower EC settings with SES during pediatric tonsillectomy significantly reduces surgical plume exposure for patients, surgeons, and operating room personnel, which is a well-known occupational health hazard.

来源:Otolaryngol Head Neck Surg,查看原文。版权归原作者所有。