薄型蜗轴电极人工耳蜗植入的即刻与1年结局
该回顾性病例系列分析了326例接受薄型蜗轴电极(SME)人工耳蜗植入的患者,术中X线识别电极尖端卷曲,76例经术后CT重建确定阶内位置。结果显示尖端卷曲率7%,阶间移位6.6%,言语感知在12个月时有显著改善,但保留听力仅20%可维持。
要点
- SME植入可实现可靠的鼓阶插入和一致的蜗轴周围位置。
- 尖端卷曲率初期较高,随病例量和鞘设计改进而下降。
- 长期言语表现与其他人工耳蜗相当,但听力保留不恒定。
核心结论
- SME人工耳蜗植入提供可靠的鼓阶插入和一致的蜗轴周围位置。
- 言语表现与其他人工耳蜗相当,但听力保留不恒定。
研究设计与样本
- 连续病例系列 with 病历回顾, tertiary referral academic center。
- 2016年5月至2018年8月共326例人工耳蜗植入。
主要发现
- 326例中23例(7%)出现尖端卷曲。
- 76例术后CT重建中5例(6.6%)发生阶间移位。
- 177例符合言语感知评估标准,12个月时辅音-核-辅音词得分较术前平均提高43.7(95% CI, 39.8-47.6)。
- AzBio安静环境得分平均提高49.7(95% CI, 44.9-54.4),噪声环境(+10-dB信噪比)得分平均提高29.9(95% CI, 25.2-34.7)。
- 61例可保留听力(低频纯音平均≤80 dB)患者中,12例(20%)在1年时保留听力。
关键数值与推荐
- 听力保留定义为250和500 Hz低频纯音平均≤80 dB。
- 言语测试在3、6、12个月进行,使用辅音-核-辅音词和AzBio句子(安静和噪声+10-dB信噪比)。
临床怎么用
- 术中X线可用于识别电极尖端卷曲,术后CT可评估阶内位置。
- SME植入后言语改善显著,但需告知患者听力保留可能性有限。
局限性
- 回顾性病例系列设计。
- 言语感知评估仅纳入177例子集。
- 听力保留分析基于61例可保留听力患者。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To explore the immediate and 1-year outcomes of patients who underwent implantation with the slim modiolar electrode (SME). STUDY DESIGN: Consecutive case series with chart review. SETTING: Tertiary referral academic center. SUBJECT AND METHODS: Between May 2016 and August 2018, a total of 326 cochlear implantations (CIs) were performed. Intraoperative x-rays were performed in all cases to identify tip rollovers. Scalar location was identified for 76 CIs that had postoperative computed tomography reconstructions. Speech outcomes were measured at 3, 6, and 12 months with consonant-nucleus-consonant word and AzBio sentences in quiet and noise (+10-dB signal-to-noise ratio). Preservation of hearing was defined as maintaining a low-frequency pure tone average ≤80 dB at 250 and 500 Hz. RESULTS: Among 326 CIs, 23 (7%) had tip rollovers. Postoperative reconstructions revealed 5 of 76 (6.6%) scalar translocations. A subset of 177 cases met criteria for evaluation of speech perception scores. The marginal mean differences between presurgery and 12 months for speech tests were as follows: consonant-nucleus-consonant, 43.7 (95% CI, 39.8-47.6); AzBio in quiet, 49.7 (95% CI, 44.9-54.4); and AzBio in noise, 29.9 (95% CI, 25.2-34.7). Sixty-one patients were identified with preservable hearing (low-frequency pure tone average ≤80 dB), and 12 of 61 (20%) preserved hearing at 1 year. CONCLUSION: CI with SME provides reliable scala tympani insertion in a consistent perimodiolar position. An initially increased tip rollover rate improved with case volume and sheath design improvement. For long-term outcomes, speech performance was comparable to that of other cochlear implants. While hearing preservation for the SME may be better than prior perimodiolar electrodes, consistent outcomes are unlikely.
引用
- Durakovic N, Kallogjeri D, Wick CC, et al. Immediate and 1-Year Outcomes with a Slim Modiolar Cochlear Implant Electrode Array. Otolaryngol Head Neck Surg. 2020. PMID: 32180504. doi:10.1177/0194599820907336
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。