Nucleus Hybrid S12 多中心临床试验 12 个月结果
这项单病例设计多中心试验纳入 28 例低频听力≤60 dB HL(500 Hz)且高频重度听损的受试者,植入 10 mm/10 电极 Nucleus Hybrid S12 人工耳蜗。12 个月时 85%(23/27)维持功能性残余听力,81% 的受试者 CNC 词识别和 77% 的 AzBio 噪声句识别较术前双耳助听器显著改善,SSQ 主观评分也显著提高。
要点
- 10 mm 短电极植入后 96% 受试者实现功能性听力保留,12 个月时 85% 仍维持。
- 12 个月时 81% 的受试者 CNC 词识别、77% 的 AzBio 噪声句识别较术前助听器显著改善。
- SSQ 主观评分从术前到术后 12 个月显著提高,支持声电联合刺激的临床获益。
核心结论
- 10 mm 短电极 Nucleus Hybrid S12 可实现高度残余听力保留, enabling 声电联合刺激,并改善安静与噪声下的言语识别。
研究设计与样本
- 多中心临床试验,采用单病例设计,共纳入 28 例低频听力≤60 dB HL(500 Hz)且高频重度听损的受试者,植入于较差耳。
主要发现
- 96% 的受试者实现功能性听力保留;3 个月和 6 个月时 86% 的 28 例受试者维持功能性听力;12 个月时 23/27 例(85%)维持功能性听力。
- 12 个月时,81% 的受试者日常聆听条件下 CNC 词识别较术前双耳助听器显著改善,77% 的受试者 AzBio 噪声句识别显著改善。
- SSQ 主观评分从术前到术后 12 个月显著改善。
临床怎么用
- 对于低频残余听力尚可(500 Hz ≤60 dB HL)且高频重度听损的患者,可考虑使用 10 mm 短电极人工耳蜗以保留残余听力并改善言语识别。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES/HYPOTHESIS: The use of a short 10-mm/10-electrode cochlear implant to preserve low-frequency residual hearing was investigated. This report describes the 12-month outcomes of this multicenter clinical trial. STUDY DESIGN: Single-subject design. METHODS: Twenty-eight subjects with low-frequency hearing at or better than 60 dB HL at 500 Hz and severe high-frequency hearing loss were implanted with a Nucleus Hybrid S12 implant in their poorer ear. Speech perception in quiet using Consonant-Nucleus-Consonant (CNC) words and sentences in noise using AzBio sentences was collected pre- and postoperatively at 3, 6, and 12 months. Subjective reporting using the Speech, Spatial, and Qualities of Hearing Scale (SSQ) questionnaire was also collected pre- and postoperatively. RESULTS: Functional hearing preservation was accomplished in 96% of subjects. At 3 and 6 months, 86% of the 28 subjects had maintained functional hearing. By 12 months, 23 out of 27 subjects (85%) had maintained functional hearing (one subject with functional hearing at 6 months withdrew from the study prior to the 12-month visit). Speech perception results demonstrated that 81% of the participants on CNC words and 77% with AzBio sentences in noise had significant improvements using their everyday listening condition at 12 months compared to preoperative performance with bilateral hearing aids. Furthermore, preoperative to 12 months postoperative subjective ratings showed significant improvements for the SSQ. CONCLUSIONS: This study demonstrates that a high degree of hearing preservation enabling acoustic-electric hearing and improvement in speech understanding in quiet and in noise can be accomplished using a short-electrode 10-mm cochlear implant. LEVEL OF EVIDENCE: 2c Laryngoscope, 130:E548-E558, 2020.
引用
- Dunn CC, Oleson J, Parkinson A, et al. Nucleus Hybrid S12: Multicenter Clinical Trial Results. Laryngoscope. 2020. PMID: 32212342. doi:10.1002/lary.28628
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。