非接触设备采集的睡眠呼吸音预测呼吸暂停低通气指数

Prediction of Apnea-Hypopnea Index Using Sound Data Collected by a Noncontact Device.
PubMed 文献 · Otolaryngol Head Neck Surg · 2020-02-04

这项前瞻性队列研究纳入116例接受整夜多导睡眠监测的患者,用非接触空气传导麦克风记录睡眠呼吸音,经降噪和分帧提取声学特征后,以高斯过程、支持向量机、随机森林和简单线性回归预测AHI。随机森林表现最佳,相关系数0.83,平均绝对误差9.64事件/h,均方根误差13.72事件/h;其他模型相关系数0.74至0.79。

要点

  • 随机森林模型预测AHI的相关系数为0.83,平均绝对误差9.64事件/h。
  • OSA越重,AHI越易被低估;重度OSA亚组的估计性能最低。
  • 该方法或可用于OSA的初筛及治疗后随访。

核心结论

  • 利用睡眠中呼吸音数据可较好地预测AHI,随机森林模型性能最高。

研究设计与样本

  • 前瞻性队列研究,共纳入116例接受有人值守、实验室整夜多导睡眠监测的患者。

主要发现

  • 随机森林的相关系数最高为0.83,平均绝对误差为9.64事件/h,均方根误差为13.72事件/h。
  • 其他模型的相关系数在0.74至0.79之间,性能略低但相近。
  • 随着OSA严重程度增加,估计的AHI倾向于被低估;重度OSA亚组的估计性能最低,与所用模型无关。
  • 在声学特征中,总体标准差的矩面积导数与AHI相关性最高。

临床怎么用

  • 该预测模型可能不仅用于OSA的初筛,还可用于患者治疗后的随访。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To predict the apnea-hypopnea index (AHI) in patients with obstructive sleep apnea (OSA) using data from breathing sounds recorded using a noncontact device during sleep. STUDY DESIGN: Prospective cohort study. SETTING: Tertiary referral hospital. SUBJECT AND METHODS: Audio recordings during sleep were performed using an air-conduction microphone during polysomnography. Breathing sounds recorded from all sleep stages were analyzed. After noise reduction preprocessing, the audio data were segmented into 5-second windows and sound features were extracted. Estimation of AHI by regression analysis was performed using a Gaussian process, support vector machine, random forest, and simple linear regression, along with 10-fold cross-validation. RESULTS: In total, 116 patients who underwent attended, in-laboratory, full-night polysomnography were included. Overall, random forest resulted in the highest performance with the highest correlation coefficient (0.83) and least mean absolute error (9.64 events/h) and root mean squared error (13.72 events/h). Other models resulted in somewhat lower but similar performances, with correlation coefficients ranging from 0.74 to 0.79. The estimated AHI tended to be underestimated as the severity of OSA increased. Regarding bias and precision, estimation performances in the severe OSA subgroup were the lowest, regardless of the model used. Among sound features, derivative of the area methods of moments of overall standard deviation demonstrated the highest correlation with AHI. CONCLUSION: AHI was fairly predictable by using data from breathing sounds generated during sleep. The prediction model may be useful not only for prescreening but also for follow-up after treatment in patients with OSA.

来源:Otolaryngol Head Neck Surg,查看原文。版权归原作者所有。

引用

  1. Kim JW, Kim T, Shin J, et al. Prediction of Apnea-Hypopnea Index Using Sound Data Collected by a Noncontact Device. Otolaryngol Head Neck Surg. 2020. PMID: 32013710. doi:10.1177/0194599819900014

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。