鼓阶外侧壁高度对直电极人工耳蜗植入深度的影响

Effect of Scala Tympani Height on Insertion Depth of Straight Cochlear Implant Electrodes.
PubMed 文献 · Otolaryngol Head Neck Surg · 2020-02-25

该回顾性研究测量177耳术前鼓阶外侧壁高度及术后电极位置,发现39耳(22%)发生 translocation,中位植入角度443°;

要点

  • 直电极未发生 translocation 时,植入角度与 CNC 词分弱正相关。
  • translocation 多发生在约380°,超过预测深度后风险显著增加。
  • 外侧壁鼓阶高度向顶侧递减,限制深部植入并增加 translocation 风险。

核心结论

  • 直电极未发生 translocation 时植入角度与 CNC 词分直接相关。
  • translocation 通常发生在约380°,深部植入因外侧壁鼓阶高度递减而更常见。
  • 超过约580°后 translocation 风险显著增加。

研究设计与样本

  • 回顾性研究,纳入177耳直电极人工耳蜗植入患者。

主要发现

  • 39耳(22%)发生 translocation,中位植入角度443°(IQR 367°-550°)。
  • 中位 translocation 深度381°(IQR 222°-399°)。

关键数值与推荐

  • 0.5-mm电极在距外侧壁0.1 mm内无法容纳的中位深度为585°(IQR 405°-585°)。
  • 0.5-mm电极使基底膜位移≥0.1 mm的中位深度(预测 translocation 深度)为585°(IQR 518°-765°)。

临床怎么用

  • 术前可测量外侧壁鼓阶高度以评估深部植入的 translocation 风险。
  • 植入深度接近或超过约580°时需警惕 translocation 风险增加。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Studies suggest lateral wall (LW) scala tympani (ST) height decreases apically, which may limit insertion depth. No studies have investigated the relationship of LW ST height with translocation rate or location. STUDY DESIGN: Retrospective review. SETTING: Cochlear implant program at tertiary referral center. SUBJECTS AND METHODS: LW ST height was measured in preoperative images for patients with straight electrodes. Scalar location, angle of insertion depth (AID), and translocation depth were measured in postoperative images. Audiologic outcomes were tracked. RESULTS: In total, 177 ears were identified with 39 translocations (22%). Median AID was 443° (interquartile range [IQR], 367°-550°). Audiologic outcomes (126 ears) showed a small, significant correlation between consonant-nucleus-consonant (CNC) word score and AID (r = 0.20, P = .027), although correlation was insignificant if translocation occurred (r = 0.11, P = .553). Translocation did not affect CNC score (P = .335). AID was higher for translocated electrodes (503° vs 445°, P = .004). Median translocation depth was 381° (IQR, 222°-399°). Median depth at which a 0.5-mm electrode would not fit within 0.1 mm of LW was 585° (IQR, 405°-585°). Median depth at which a 0.5-mm electrode would displace the basilar membrane by ≥0.1 mm was 585° (IQR, 518°-765°); this was defined as predicted translocation depth (PTD). Translocation rate was 39% for insertions deeper than PTD and 14% for insertions shallower than PTD (P = .008). CONCLUSION: AID and CNC are directly correlated for straight electrodes when not translocated. Translocations generally occur around 380° and are more common with deeper insertions due to decreasing LW ST height. Risk of translocation increases significantly after 580°.

来源:Otolaryngol Head Neck Surg,查看原文。版权归原作者所有。