电场成像/跨阻抗测量检测人工耳蜗耳外电极:人体尸体研究
本研究在8例新鲜冷冻人体尸头中植入Advanced Bionics HiRes90K(HiFocus 1J外侧壁电极),记录接触阻抗和SCINSEV,评估其检测耳外电极的能力。结果显示,在 saline 或软组织中,自动分析工具检测两个及以上耳外电极的敏感性和特异性均为100%,量化准确率分别为84%和81%。SCINSEV有望替代CT用于耳外电极的初步筛查。
要点
- SCINSEV可检测耳外电极,即使接触阻抗与耳内电极相似。
- 自动分析工具检测≥2个耳外电极的敏感性和特异性均达100%。
- SCINSEV有望替代CT用于耳外电极的初步筛查。
核心结论
- SCINSEV(EFI)在检测耳外电极方面显示出良好潜力,即使接触阻抗与耳内电极相似。
- SCINSEV可能替代CT用于耳外电极的初步筛查。
研究设计与样本
- 8例新鲜冷冻人体尸头植入Advanced Bionics HiRes90K with HiFocus 1J外侧壁电极。
- 通过外半规管用1.0% saline冲洗耳蜗,记录完全插入和1至5个耳外电极时的接触阻抗和SCINSEV。
主要发现
- 完全插入且耳蜗冲洗后,接触阻抗范围为6.06 ± 2.99 kΩ(均值 ± 2SD)。
- 耳外电极位于空气中时接触阻抗高,但在saline或软组织中与耳内接触阻抗相似。
- SCINSEV在空气、saline和软组织中耳外电极处显示形状变化。
- 自动分析工具检测saline和软组织中两个及以上耳外电极的敏感性和特异性均为100%。
- 两个及以上耳外电极的量化在saline和软组织测量中分别正确84%和81%。
临床怎么用
- SCINSEV可用于耳外电极的检测,尤其在接触阻抗与耳内电极相似时。
- 自动检测工具可辅助检测和量化两个及以上耳外电极。
- 在手术最后阶段检测电极阵列迁移可能避免部分CI用户再次插入手术。
局限性
- 研究基于人体尸体,临床适用性需进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Extracochlear electrodes in cochlear implants (CI), defined as individual electrodes on the electrode array located outside of the cochlea, are not a rare phenomenon. The presence of extracochlear electrodes frequently goes unnoticed and could result in them being assigned stimulation frequencies that are either not delivered to, or stimulating neurons that overlap with intracochlear electrodes, potentially reducing performance. The current gold-standard for detection of extracochlear electrodes is computed tomography (CT), which is time-intensive, costly and involves radiation. It is hypothesized that a collection of Stimulation-Current-Induced Non-Stimulating Electrode Voltage recordings (SCINSEVs), commonly referred to as "transimpedance measurements (TIMs)" or electric field imaging (EFI), could be utilized to detect extracochlear electrodes even when contact impedances are low. An automated analysis tool is introduced for detection and quantification of extracochlear electrodes. DESIGN: Eight fresh-frozen human cadaveric heads were implanted with the Advanced Bionics HiRes90K with a HiFocus 1J lateral-wall electrode. The cochlea was flushed with 1.0% saline through the lateral semicircular canal. Contact impedances and SCINSEVs were recorded for complete insertion and for 1 to 5 extracochlear electrodes. Measured conditions included: air in the middle ear (to simulate electrodes situated in the middle ear), 1.0% saline in the middle ear (to simulate intraoperative conditions with saline or blood in the middle ear), and soft tissue (temporal muscle) wrapped around the extracochlear electrodes (to simulate postoperative soft-tissue encapsulation of the electrodes). Intraoperative SCINSEVs from patients were collected, for clinical purposes during slow insertion of the electrode array, as well as from a patient postoperatively with known extracochlear electrodes. RESULTS: Full insertion of the cochlear implant in the fresh-frozen human cadaveric heads with a flushed cochlea resulted in contact impedances in the range of 6.06 ± 2.99 kΩ (mean ± 2SD). Contact impedances were high when the extracochlear electrodes were located in air, but remained similar to intracochlear contact impedances when in saline or soft tissue. SCINSEVs showed a change in shape for the extracochlear electrodes in air, saline, and soft tissue. The automated analysis tool showed a specificity and sensitivity of 100% for detection of two or more extracochlear electrodes in saline and soft tissue. The quantification of two or more extracochlear electrodes was correct for 84% and 81% of the saline and soft tissue measurements, respectively. CONCLUSIONS: Our analysis of SCINSEVs (specifically the EFIs from this manufacturer) shows good potential as a detection tool for extracochlear electrodes, even when contact impedances remain similar to intracochlear values. SCINSEVs could potentially replace CT in the initial screening for extracochlear electrodes. Detecting migration of the electrode array during the final stages of surgery could potentially prevent re-insertion surgery for some CI users. The automated detection tool could assist in detection and quantification of two or more extracochlear electrodes.
引用
- de Rijk SR, Tam YC, Carlyon RP, et al. Detection of Extracochlear Electrodes in Cochlear Implants with Electric Field Imaging/Transimpedance Measurements: A Human Cadaver Study. Ear Hear. 2020. PMID: 31923041. doi:10.1097/AUD.0000000000000837