幻影刺激在人工耳蜗使用者中改变音高感知的有效性
研究比较了多种幻影电极配置在人工耳蜗使用者中产生的兴奋位置偏移,通过音高匹配实验测量。
要点
- 幻影刺激可显著向蜗顶偏移兴奋位置,平均偏移0.39-0.76个电极触点。
- 最佳配置个体间差异大,需个体化优化幻影电极设置。
- 蜗顶和侧壁电极触点的偏移更大,刺激水平不影响偏移。
核心结论
- 幻影刺激能产生显著的兴奋位置偏移,尤其在电极阵列的蜗顶部分。
- 产生最大偏移的幻影配置因人而异,因此幻影电极设置应个体化以优化每位使用者。
- 对音质的实际附加价值需通过家庭试用确定。
研究设计与样本
- 包含四个实验:音高辨别实验确定电极触点,响度平衡实验确定T级和M级,音高匹配实验估计兴奋位置偏移,后续实验更精确测定偏移。
- 测试了五种幻影配置:对称双相脉冲SBC0.5、SBC0.6、SBC0.7、SBC0.8(σ分别为0.5、0.6、0.7、0.8)和伪单相脉冲PSA0.75(σ=0.75)。
主要发现
- 从20个电极触点获得可靠数据。
- 平均偏移分别为:SBC0.5为0.39、SBC0.6为0.53、SBC0.7为0.64、SBC0.8为0.76、PSA0.75为0.53个电极触点(向蜗顶)。
- 仅纳入每个电极触点最佳配置时,平均兴奋位置偏移为0.92个电极触点(范围0.25至2.0)。
- PSA0.75在蜗顶与SBC配置结果相当,但在蜗底未导致显著偏移。
- 兴奋位置偏移在蜗顶和侧壁电极触点显著更大;刺激水平不影响偏移。
临床怎么用
- 幻影刺激可用于提供更低的音高感知,无需加深电极插入。
- 临床应个体化设置幻影电极参数以优化每位人工耳蜗使用者。
- 需通过家庭试用验证幻影刺激对音质的实际增益。
局限性
- 研究为实验性测量,未评估长期音质受益。
- 结论需通过家庭试用进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Phantom electrode stimulation was developed for cochlear implant (CI) systems to provide a lower pitch percept by stimulating more apical regions of the cochlea, without inserting the electrode array deeper into the cochlea. Phantom stimulation involves simultaneously stimulating a primary and a compensating electrode with opposite polarity, thereby shifting the electrical field toward the apex and eliciting a lower pitch percept. The current study compared the effect sizes (in shifts of place of excitation) of multiple phantom configurations by matching the perceived pitch with phantom stimulation to that perceived with monopolar stimulation. Additionally, the effects of electrode location, type of electrode array, and stimulus level on the perceived pitch were investigated. DESIGN: Fifteen adult advanced bionics CI users participated in this study, which included four experiments to eventually measure the shifts in place of excitation with five different phantom configurations. The proportions of current delivered to the compensating electrode, expressed as σ, were 0.5, 0.6, 0.7, and 0.8 for the symmetrical biphasic pulses (SBC0.5, SBC0.6, SBC0.7, and SBC0.8) and 0.75 for the pseudomonophasic pulse shape (PSA0.75). A pitch discrimination experiment was first completed to determine which basal and apical electrode contacts should be used for the subsequent experiments. An extensive loudness balancing experiment followed where both the threshold level (T-level) and most comfortable level (M-level) were determined to enable testing at multiple levels of the dynamic range. A pitch matching experiment was then performed to estimate the shift in place of excitation at the chosen electrode contacts. These rough shifts were then used in the subsequent experiment, where the shifts in place of excitation were determined more accurately. RESULTS: Reliable data were obtained from 20 electrode contacts. The average shifts were 0.39, 0.53, 0.64, 0.76, and 0.53 electrode contacts toward the apex for SBC0.5, SBC0.6, SBC0.7, SBC0.8, and PSA0.75, respectively. When only the best configurations per electrode contact were included, the average shift in place of excitation was 0.92 electrode contacts (range: 0.25 to 2.0). While PSA0.75 leads to equal results as the SBC configurations in the apex, it did not result in a significant shift at the base. The shift in place of excitation was significantly larger at the apex and with lateral wall electrode contacts. The stimulus level did not affect the shift. CONCLUSIONS: Phantom stimulation results in significant shifts in place of excitation, especially at the apical part of the electrode array. The phantom configuration that leads to the largest shift in place of excitation differs between subjects. Therefore, the settings of the phantom electrode should be individualized so that the phantom stimulation is optimized for each CI user. The real added value to the sound quality needs to be established in a take-home trial.
引用
- de Jong MAM, Briaire JJ, Biesheuvel JD, et al. Effectiveness of Phantom Stimulation in Shifting the Pitch Percept in Cochlear Implant Users. Ear Hear. 2020. PMID: 31977727. doi:10.1097/AUD.0000000000000845
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。