文献速递咽喉与嗓音

犬全身麻醉下气流驱动声带振动个体化评估

Aerodynamically driven phonation of individual vocal folds under general anesthesia in canines.
PubMed 文献 · Laryngoscope · 2019-10-11

本研究在犬模型中测试了气流驱动声带振动器(AVFD)在全身麻醉显微喉镜下的效果,共评估26条声带,包括4条电灼瘢痕声带。结果显示所有声带均成功驱动振动,15mm宽AVFD最适配声带,接触压力与声学输出、驱动压力和振动模式相关,为声外科决策提供信息。

要点

  • AVFD可在全麻显微喉镜下驱动单侧声带振动,无需患者配合。
  • 15mm宽AVFD最适配声带前后径,接触压力与声学输出相关。
  • 瘢痕声带仍可被驱动振动,有助于评估常规清醒发音不可见的信息。

核心结论

  • AVFD可在全麻显微喉镜下提供单侧声带振动特征,可能辅助声外科决策。

研究设计与样本

  • 动物模型研究,在犬显微喉镜下测试AVFD,共评估26条声带(瘢痕4条,对照22条)。

主要发现

  • 所有26条声带均成功驱动振动,包括瘢痕阻止常规双侧振动者。
  • 15mm宽AVFD最适配肌腱膜性声带前后径。
  • 声带接触压力与声学输出、驱动压力和可见振动模式相关。

关键数值与推荐

  • AVFD为手持器械,经其轴杆向声门下输送气流驱动声带振动。
  • 采用原始设计或三种尺寸的医用级聚合物3D打印版本。
  • 内置高保真压力传感器测量声带接触压力。

临床怎么用

  • 可在全麻显微喉镜下放大观察单侧声带黏膜波功能。
  • 无需患者参与,可能提供常规清醒发音不可见的信息。
  • 适用于瘢痕声带评估,辅助声外科决策。

局限性

  • 研究为犬动物模型,样本量有限。
  • 证据等级为NA。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES/HYPOTHESIS: We previously developed an instrument called the Aerodynamic Vocal Fold Driver (AVFD) for intraoperative magnified assessment of vocal fold (VF) vibration during microlaryngoscopy under general anesthesia. Excised larynx testing showed that the AVFD could provide useful information about the vibratory characteristics of each VF independently. The present investigation expands those findings by testing new iterations of the AVFD during microlaryngoscopy in the canine model. STUDY DESIGN: Animal model. METHODS: The AVFD is a handheld instrument that is positioned to contact the phonatory mucosa of either VF during microlaryngoscopy. Airflow delivered through the AVFD shaft to the subglottis drives the VF into phonation-like vibration, which enables magnified observation of mucosal-wave function with stroboscopy or high-speed video. AVFD-driven phonation was tested intraoperatively (n = 26 VFs) using either the original instrument design or smaller and larger versions three-dimensionally printed from a medical grade polymer. A high-fidelity pressure sensor embedded within the AVFD measured VF contact pressure. Characteristics of individual VF phonation were compared with typical two-fold phonation and compared for VFs scarred by electrocautery (n = 4) versus controls (n = 22). RESULTS: Phonation was successful in all 26 VFs, even when scar prevented conventional bilateral phonation. The 15-mm-wide AVFD fits best within the anteroposterior dimension of the musculo-membranous VF, and VF contact pressure correlated with acoustic output, driving pressures, and visible modes of vibration. CONCLUSIONS: The AVFD can reveal magnified vibratory characteristics of individual VFs during microlaryngoscopy (e.g., without needing patient participation), potentially providing information that is not apparent or available during conventional awake phonation, which might facilitate phonosurgical decision making. LEVEL OF EVIDENCE: NA Laryngoscope, 130: 1980-1988, 2020.

来源:Laryngoscope,查看原文。版权归原作者所有。

引用

  1. Heaton JT, Kobler JB, Ottensmeyer MP, et al. Aerodynamically driven phonation of individual vocal folds under general anesthesia in canines. Laryngoscope. 2019. PMID: 31603575. doi:10.1002/lary.28333

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。