文献速递鼻科

球囊鼻窦扩张术相关不良事件:MAUDE数据库分析

Adverse Events Associated with Balloon Sinuplasty: A MAUDE Database Analysis.
PubMed 文献 · Otolaryngol Head Neck Surg · 2019-10-22

该研究基于FDA的MAUDE数据库(2008-2018年)回顾性分析球囊鼻窦扩张术相关不良事件,共纳入3家主要制造商的208份报告、211例不良事件。患者相关事件占48.3%,器械相关占47.9%,包装相关占3.8%;最常见的器械并发症为导引导管故障(39.6%)、球囊故障(38.6%)和导航不精确(17.8%),最常见的患者并发症为脑脊液漏(36.3%)、眼部肿胀(29.4%)和鼻出血(11.8%)。

要点

  • 球囊鼻窦扩张术最常见不良事件为球囊故障、导引导管故障、脑脊液漏和显著眼部肿胀。
  • 报告4例围手术期死亡,但与技术并发症无明确关联。

核心结论

  • 球囊鼻窦扩张术最常见的不良事件包括球囊故障、导引导管故障、脑脊液漏和显著眼部肿胀。
  • 医疗提供者在获取患者对球囊鼻窦扩张术的知情同意时应讨论这些可能的并发症。

研究设计与样本

  • 回顾性横断面分析,检索FDA的MAUDE数据库(2008-2018年)中3家主要制造商(Acclarent、Entellus、Medtronic)球囊鼻窦扩张装置的所有不良事件报告。
  • 共纳入208份报告,涉及211例不良事件。

主要发现

  • 患者相关事件102例(48.3%),器械相关事件101例(47.9%),包装相关事件8例(3.8%)。
  • 报告4例围手术期死亡,但与技术并发症无明确关联。
  • 器械相关并发症中最常见为导引导管故障(39.6%)、球囊故障(38.6%)和导航不精确(17.8%)。
  • 患者相关并发症中最常见为脑脊液漏(36.3%)、眼部肿胀(29.4%)和鼻出血(11.8%)。

临床怎么用

  • 临床在获取球囊鼻窦扩张术知情同意时,应向患者说明球囊故障、导引导管故障、脑脊液漏和显著眼部肿胀等可能并发症。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Balloon sinuplasty utilization has increased significantly since its introduction over a decade ago. However, the most common associated complications are still unknown. The objective of this study was to analyze adverse events related to balloon sinuplasty. STUDY DESIGN: Retrospective cross-sectional analysis. SETTING: Food and Drug Administration's MAUDE database (Manufacturer and User Facility Device Experience; 2008-2018). SUBJECTS AND METHODS: The MAUDE database was searched for all reports on adverse events involving balloon sinuplasty devices from the 3 leading manufacturers: Acclarent, Entellus, and Medtronic. Reported events were reviewed and categorized. RESULTS: During the study period, there were 211 adverse events from 208 reports divided into the following categories: patient related (n = 102, 48.3%), device related (n = 101, 47.9%), and packaging related (n = 8, 3.8%). Four periprocedural deaths were reported but were not clearly associated with technical complications. The most common device-related complications were guide catheter malfunction (39.6%), balloon malfunction (38.6%), and imprecise navigation (17.8%). The most common patient-related complications were cerebrospinal fluid leak (36.3%), eye swelling (29.4%), and epistaxis (11.8%). A lateral canthotomy was performed in 30.0% of eye-swelling complications. Sixty percent of eye complications occurred during balloon dilation of the maxillary sinus. The years 2014 (n = 48) and 2012 (n = 32) had the highest number of adverse events reported as compared with all other years. CONCLUSION: The most common adverse events associated with balloon sinuplasty include balloon malfunction, guide catheter malfunction, cerebrospinal fluid leak, and significant eye swelling. Health care providers should discuss these possible complications when consenting patients for balloon sinuplasty.

来源:Otolaryngol Head Neck Surg,查看原文。版权归原作者所有。

引用

  1. Hur K, Ge M, Kim J, et al. Adverse Events Associated with Balloon Sinuplasty: A MAUDE Database Analysis. Otolaryngol Head Neck Surg. 2019. PMID: 31638866. doi:10.1177/0194599819884902