文献速递耳科与耳神经

电声刺激交互影响同侧EAS用户的言语感知

Interaction Between Electric and Acoustic Stimulation Influences Speech Perception in Ipsilateral EAS Users.
PubMed 文献 · Ear Hear · 2020-01-01

心理声学实验证实电刺激可掩蔽声刺激,最大掩蔽约一个倍频程;overlap拟合在稳态噪声中言语接收显著下降,UNMASKfit在降低电刺激的同时获得相似言语阈值并增强声学获益。

要点

  • 电刺激可显著提高同侧声探测阈值,最大掩蔽位于电声频率差约一个倍频程处。
  • overlap拟合在稳态噪声中言语接收显著下降,掩蔽越强下降越明显。
  • UNMASKfit在减少电刺激的同时获得相似言语阈值并增强声学获益。

核心结论

  • 心理声学实验证实同侧电声刺激存在交互掩蔽。
  • overlap拟合在稳态噪声中言语接收较差,尤其在强掩蔽受试者中。
  • UNMASKfit策略获得相似言语阈值并增强声学获益,同时减少电刺激。

研究设计与样本

  • 三种拟合策略各使用4周:meet、overlap和UNMASKfit。
  • 采用心理声学掩蔽程序测量电掩蔽对声阈值的影响,言语接收用奥尔登堡矩阵句测试在噪声中评估。

主要发现

  • 电掩蔽存在时声探测音行为阈值显著升高。
  • 电声刺激位置频率差约一个倍频程时掩蔽最大。
  • 言语接收得分和掩蔽强度与残余听力相关。
  • overlap拟合策略下言语接收显著降低。
  • 电声刺激显著优于单独电刺激,UNMASKfit有更大获益趋势。

临床怎么用

  • 对于同侧残余听力用户,overlap拟合可能因掩蔽损害稳态噪声下言语接收。
  • UNMASKfit可在减少电刺激的同时维持言语接收并增强声学获益。
  • 可考虑在首次拟合时即采用UNMASKfit以延长适应时间。

局限性

  • 研究周期较短,长期效果未知。
  • 结果基于特定电极(MED-EL Flex)和拟合策略,外推需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The aim of this study was to determine electric-acoustic masking in cochlear implant users with ipsilateral residual hearing and different electrode insertion depths and to investigate the influence on speech reception. The effects of different fitting strategies-meet, overlap, and a newly developed masking adjusted fitting (UNMASKfit)-on speech reception are compared. If electric-acoustic masking has a detrimental effect on speech reception, the individualized UNMASKfit map might be able to reduce masking and thereby enhance speech reception. DESIGN: Fifteen experienced MED-EL Flex electrode recipients with ipsilateral residual hearing participated in a crosssover design study using three fitting strategies for 4 weeks each. The following strategies were compared: (1) a meet fitting, dividing the frequency range between electric and acoustic stimulation, (2) an overlap fitting, delivering part of the frequency range both acoustically and electrically, and (3) the UNMASKfit, reducing the electric stimulation according to the individual electric-on-acoustic masking strength. A psychoacoustic masking procedure was used to measure the changes in acoustic thresholds due to the presence of electric maskers. Speech reception was measured in noise with the Oldenburg Matrix Sentence test. RESULTS: Behavioral thresholds of acoustic probe tones were significantly elevated in the presence of electric maskers. A maximum of masking was observed when the difference in location between the electric and acoustic stimulation was around one octave in place frequency. Speech reception scores and strength of masking showed a dependency on residual hearing, and speech reception was significantly reduced in the overlap fitting strategy. Electric- acoustic stimulation significantly improved speech reception over electric stimulation alone, with a tendency toward a larger benefit with the UNMASKfit map. In addition, masking was significantly inversely correlated to the speech reception performance difference between the overlap and the meet fitting. CONCLUSIONS: (1) This study confirmed the interaction between ipsilateral electric and acoustic stimulation in a psychoacoustic masking experiment. (2) The overlap fitting yielded poorer speech reception performance in stationary noise especially in subjects with strong masking. (3) The newly developed UNMASKfit strategy yielded similar speech reception thresholds with an enhanced acoustic benefit, while at the same time reducing the electric stimulation. This could be beneficial in the long-term if applied as a standard fitting, as hair cells are exposed to less possibly adverse electric stimulation. In this study, the UNMASKfit allowed the participants a better use of their natural hearing even after 1 month of adaptation. It might be feasible to transfer these results to the clinic, by fitting patients with the UNMASKfit upon their first fitting appointment, so that longer adaptation times can further improve speech reception.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Imsiecke M, Krüger B, Büchner A, et al. Interaction Between Electric and Acoustic Stimulation Influences Speech Perception in Ipsilateral EAS Users. Ear Hear. 2020. PMID: 31592902. doi:10.1097/AUD.0000000000000807

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。