文献速递耳科与耳神经

通过频谱掩蔽释放提升人工耳蜗风噪下表现:多麦克风多通道策略

Improving Cochlear Implant Performance in the Wind Through Spectral Masking Release: A Multi-microphone and Multichannel Strategy.
PubMed 文献 · Ear Hear · 2020-01-01

本研究比较了全向麦克风模式与降低低频增益两种常见风噪处理策略对人工耳蜗使用者言语理解和音质偏好的影响。研究还检验了一种新策略,即在每个频带选择风噪较低的麦克风模式。结果表明,组合全向与指向性麦克风的策略(COMBO)在言语识别得分和音质偏好上均最优,而全向模式和低频增益降低策略效果较差。

要点

  • COMBO策略在四种麦克风模式中言语识别得分最高且最受偏好。
  • 全向麦克风模式下言语识别得分 consistently 低于COMBO,且常被评较差。
  • 降低低频增益的ADM策略得分等于或低于ADM,因同时衰减了语音低频能量。

核心结论

  • 采用全向麦克风或降低低频增益均未比COMBO策略产生更高的言语识别得分或音质评分。
  • 在不同频带选择风噪较低的麦克风模式可提供频谱掩蔽释放,是更有效的风噪抑制策略。

研究设计与样本

  • 使用多麦克风模式耳背式数字助听器在声学处理风洞中录制测试材料,风速为4.5、9.0和13.5 m/s,角度0°至360°步进10°。
  • 选取9.0 m/s风速下90°和300°头角的两段风噪样本与相同设置录制的语句混合,由人工耳蜗参与者进行言语识别和音质偏好评估。

主要发现

  • COMBO在四种麦克风模式中言语识别得分最高,且最常被偏好,可能源于频谱掩蔽的减少。
  • 采用低频滚降的ADM言语识别得分等于或低于ADM,因为增益降低不仅减少了风噪水平也减少了语音低频能量。
  • OMNI的言语识别得分始终低于COMBO,且常被评为不如其他麦克风模式。

临床怎么用

  • 在风噪环境中,不应常规切换到全向麦克风模式。
  • 当存在语音时,应补偿一阶指向性麦克风自然的6 dB/倍频程低频滚降。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Adopting the omnidirectional microphone (OMNI) mode and reducing low-frequency gain are the two most commonly used wind noise reduction strategies in hearing devices. The objective of this study was to compare the effectiveness of these two strategies on cochlear implant users' speech-understanding abilities and perceived sound quality in wind noise. We also examined the effectiveness of a new strategy that adopts the microphone mode with lower wind noise level in each frequency channel. DESIGN: A behind-the-ear digital hearing aid with multiple microphone modes was used to record testing materials for cochlear implant participants. It was adjusted to have linear amplification, flat frequency response when worn on a Knowles Electronic Manikin for Acoustic Research to remove the head-related transfer function of the manikin and to mimic typical microphone characteristics of hearing devices. Recordings of wind noise samples and hearing-in-noise test sentences were made when the hearing aid was programmed to four microphone modes, namely (1) OMNI; (2) adaptive directional microphone (ADM); (3) ADM with low-frequency roll-off; and (4) a combination of omnidirectional and directional microphone (COMBO). Wind noise samples were recorded in an acoustically treated wind tunnel from 0° to 360° in 10° increment at a wind velocity of 4.5, 9.0, and 13.5 m/s when the hearing aid was worn on the manikin. Two wind noise samples recorded at 90° and 300° head angles at the wind velocity of 9.0 m/s were chosen to take advantage of the spectral masking release effects of COMBO. The samples were then mixed with the sentences recorded using identical settings. Cochlear implant participants listened to the speech-in-wind testing materials and they repeated the sentences and compared overall sound quality preferences of different microphone modes using a paired-comparison categorical rating paradigm. The participants also rated their preferences of wind-only samples. RESULTS: COMBO yielded the highest speech recognition scores among the four microphone modes, and it was also preferred the most often, likely due to the reduction of spectral masking. The speech recognition scores generated using ADM with low-frequency roll-off were either equal to or lower than those obtained using ADM because gain reduction decreased not only the level of wind noise but also the low-frequency energy of speech. OMNI consistently yielded speech recognition scores lower than COMBO, and it was often rated as less preferable than other microphone modes, suggesting the conventional strategy to switch to the omnidirectional mode in the wind was undesirable. CONCLUSIONS: Neither adopting an OMNI nor reducing low-frequency gain generated higher speech recognition scores or higher sound quality ratings than COMBO. Adopting the microphone with lower wind noise level in different frequency channels can provide spectral masking release, and it is a more effective wind noise reduction strategy. The natural 6 dB/octave low-frequency roll-off of first-order directional microphones should be compensated when speech is present. Signal detection and decision rules for wind noise reduction applications are discussed in hearing devices with and without binaural transmission capability.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Chung K. Improving Cochlear Implant Performance in the Wind Through Spectral Masking Release: A Multi-microphone and Multichannel Strategy. Ear Hear. 2020. PMID: 31425361. doi:10.1097/AUD.0000000000000766

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。