耳蜗植入钻孔术式的人耳病理学研究
该回顾性耳病理学研究分析5例因耳蜗纤维骨化行钻孔术式植入耳蜗电极的颞骨标本,发现多例钻孔不精确造成关键结构损伤,2例损伤蜗轴并伴神经元显著减少或缺失,2例因基底转持续阻塞导致电极阵列误入前庭。
要点
- 5例颞骨标本中4例行基底转钻孔,1例行根治性耳蜗钻孔。
- 2例损伤蜗轴,相应区域神经元显著减少或缺失。
- 2例因钻孔不充分或基底转广泛骨化,电极阵列误入前庭。
核心结论
- 耳病理学揭示耳蜗纤维骨化病例钻孔术式面临挑战,钻孔路径不精确可损伤蜗轴及邻近神经元,并可能导致电极阵列误入前庭,二者均可能 contributing 于较差的听力结局。
研究设计与样本
- 回顾性耳病理学研究,纳入因术前耳蜗内纤维骨化需行钻孔术式插入耳蜗电极的颞骨标本5例。
主要发现
- 5例颞骨标本中4例行基底转钻孔,1例行根治性耳蜗钻孔。
- 多例颞骨标本钻孔不精确,造成关键结构损伤。
- 2例颞骨标本显示蜗轴损伤,相应区域神经元群体显著减少甚至缺失。
- 2例因钻孔不充分或基底转广泛骨化导致基底转持续阻塞,电极阵列被置于耳蜗外的前庭内。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Human otopathology following drill-out procedures for cochlear implantation (CI) in cases with labyrinthitis ossificans (LO) has not been previously described. This study uses the high sensitivity of histopathology to (1) evaluate surgical drill-out technique with associated intracochlear findings and (2) quantify spiral ganglion neuron populations in a series of patients with LO who underwent CI. STUDY DESIGN: Retrospective otopathology study. SETTING: Otopathology laboratory. SUBJECTS AND METHODS: Temporal bone (TB) specimens from cases with evidence of preoperative intracochlear fibroossification that required a drill-out procedure for CI electrode array insertion were included. All cases were histopathologically evaluated and 3-dimensional reconstructions of the cochleae were performed to interpret drilling paths and electrode trajectories. RESULTS: Five TB specimens were identified, of which 4 underwent drill-out of the basal turn of the cochlea and 1 underwent a radical cochlear drill-out. In multiple TBs, drilling was imprecise with resultant damage to essential structures. Two TBs showed injury to the modiolus, which was associated with substantially decreased or even absent neuronal populations within these areas. In addition, 2 cases with inadequate drill-out or extensive LO of the basal turn resulted in extracochlear placement of electrode arrays into the vestibule due to persistent obstruction within the basal turn. CONCLUSION: Otopathology highlights the challenges of drill-out procedures in cases of LO. Imprecise drilling paths, due to distortion of normal cochlear anatomy, risk injury to the modiolus and adjacent neurons as well as extracochlear placement of electrode arrays, both of which may contribute to poorer hearing outcomes.
引用
- Trakimas DR, Ishai R, Kozin ED, et al. Human Otopathology of Cochlear Implant Drill-out Procedures. Otolaryngol Head Neck Surg. 2019. PMID: 31060442. doi:10.1177/0194599819847636
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。