孤立性腭后塌陷患者舌下神经上气道刺激的疗效
回顾性分析82例接受上气道刺激(UAS)装置植入的睡眠呼吸障碍患者,其中25例为孤立性腭后塌陷,57例为其他塌陷模式;两组术后呼吸暂停低通气指数(AHI)和最低血氧饱和度(NOS)均显著改善,且组间无显著差异。
要点
- 孤立性腭后塌陷患者术后AHI为5.7,其他塌陷模式患者为3.9,组间无显著差异(P=.888)。
- 孤立性腭后塌陷患者术后NOS为92%,其他患者为91%,组间无显著差异(P=.402)。
- 所有患者术后AHI和NOS均较术前显著改善。
核心结论
- 所有患者在接受UAS装置植入后均表现出显著改善。
- 孤立性腭后塌陷患者在AHI和NOS方面的改善与其他类型塌陷患者相似。
研究设计与样本
- 回顾性综述,单机构三级学术医疗中心。
- 91例患者植入UAS装置,82例符合分析纳入标准;其中25例为孤立性腭后塌陷,57例为其他塌陷模式。
主要发现
- 所有患者术前平均AHI为38.7(95% CI,35.0-42.4),术后改善至4.5(95% CI,2.3-6.6)。
- 所有患者术前平均NOS为78%(95% CI,75%-80%),术后改善至91%(95% CI,91%-92%)。
- 两组术前人口学特征、AHI或NOS无显著差异。
- 术后AHI:孤立性腭后塌陷组为5.7(95% CI,0.57-10.8),其他患者组为3.9(95% CI,1.7-6.1),P=.888。
- 术后NOS:孤立性腭后塌陷组为92%(95% CI,90%-94%),其他患者组为91%(95% CI,90%-92%),P=.402。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To examine whether patients with isolated retropalatal collapse perform as well as others following implantation with an upper airway stimulation (UAS) device. STUDY DESIGN: Retrospective review. SETTING: Single-institution tertiary academic care medical center. SUBJECTS AND METHODS: Following drug-induced sleep endoscopy, subjects who met inclusion criteria for implantation with a UAS device received an implant per industry standard. Subjects with isolated retropalatal collapse were compared with those having other patterns of collapse. Outcome measures included apnea-hypopnea index (AHI) and nadir oxyhemoglobin saturation (NOS). RESULTS: Ninety-one patients were implanted during the duration of the study, and 82 met inclusion criteria for analysis. Twenty-five had isolated retropalatal collapse, while the remaining 57 had other patterns of collapse on drug-induced sleep endoscopy. For all patients, mean preoperative AHI and NOS were 38.7 (95% CI, 35.0-42.4) and 78% (95% CI, 75%-80%), respectively; these improved postoperatively to 4.5 (95% CI, 2.3-6.6) and 91% (95% CI, 91%-92%). There was no significant preoperative difference between groups with regard to demographics, AHI, or NOS. Group comparison showed postoperative AHI to be 5.7 (95% CI, 0.57-10.8) for patients with isolated retropalatal collapse and 3.9 (95% CI, 1.7-6.1) for other patients ( P = .888). Postoperative NOS was 92% (95% CI, 90%-94%) among patients with isolated retropalatal collapse and 91% (95% CI, 90%-92%) for others ( P = .402). CONCLUSIONS: All patients showed significant improvement following implantation with UAS. Patients with isolated retropalatal collapse showed similar improvement to other types of collapse with regard to AHI and NOS.
引用
- Mahmoud AF, Thaler ER. Outcomes of Hypoglossal Nerve Upper Airway Stimulation among Patients with Isolated Retropalatal Collapse. Otolaryngol Head Neck Surg. 2019. PMID: 30857472. doi:10.1177/0194599819835186
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。