声带创面闭合手术粘合剂的实验研究

Investigation of surgical adhesives for vocal fold wound closure.
PubMed 文献 · Laryngoscope · 2019-03-18

本研究在150个猪喉上测试三种手术粘合剂(Glubran 2、BioGlue、Tisseel)用于声带创面闭合的剪切与拉伸粘合强度,并用中性红摄取法检测其对永生化人声带成纤维细胞的细胞毒性。结果显示粘合强度与毒性之间存在明显权衡:Glubran 2粘合最强但毒性最高,Tisseel粘合最弱但毒性最低。

要点

  • 三种粘合剂剪切粘合强度:Glubran 2为68.79±13.29 kPa,BioGlue为40.92±17.94 kPa,Tisseel为13.86±5.03 kPa。
  • 三种粘合剂拉伸粘合强度:Glubran 2为46.67±12.13 kPa,BioGlue为34.27±12.59 kPa,Tisseel为10.70±6.42 kPa。
  • 声带细胞活力:Tisseel提取物为99.27%,BioGlue为43.05%,Glubran 2为1.79%。

核心结论

  • 粘合强度与毒性之间存在明显权衡,剪切或拉伸最大失败强度由强到弱依次为Glubran 2、BioGlue、Tisseel,Tisseel毒性最低而Glubran 2毒性最高。

研究设计与样本

  • 在150个猪喉上进行剪切和拉伸测试,并用中性红摄取法检测粘合剂对永生化人声带成纤维细胞的细胞毒性。

主要发现

  • Tisseel、BioGlue和Glubran 2的平均剪切粘合强度分别为13.86±5.03 kPa、40.92±17.94 kPa和68.79±13.29 kPa。
  • Tisseel、BioGlue和Glubran 2的拉伸粘合强度分别为10.70±6.42 kPa、34.27±12.59 kPa和46.67±12.13 kPa。
  • Tisseel、BioGlue和Glubran 2提取物中的声带细胞活力分别为99.27%、43.05%和1.79%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Surgical adhesives are increasingly used for vocal fold microsurgery to assist wound closure and reduce the risks of scar formation. Currently used vocal fold adhesives such as fibrin glue, however, have thus far not been found to promote wound closure or reduce scarring. The objectives of this study were to investigate the mechanical strength and the cytotoxicity of three commercially available adhesives (Glubran 2, GEM, Viareggio, Italy; BioGlue, CryoLife, Kennesaw, GA; and Tisseel, Baxter Healthcare, Deerfield, IL) for vocal fold wound closure. METHODS: Shear and tension tests were performed on 150 porcine larynges. The cytotoxicity of the adhesives to immortalized human vocal fold fibroblasts was investigated using neutral red uptake assays. RESULTS: The average shear adhesive strength for Tisseel, BioGlue, and Glubran 2 was 13.86 ± 5.03 kilopascal (kPa), 40.92 ± 17.94 kPa, and 68.79 ± 13.29 kPa, respectively. The tensile adhesive strength for Tisseel, BioGlue, and Glubran 2 was 10.70 ± 6.42 kPa, 34.27 ± 12.59 kPa, and 46.67 ± 12.13 kPa, respectively. The vocal fold cell viabilities in extracts of Tisseel, BioGlue, and Glubran 2 were 99.27%, 43.05%, and 1.79%, respectively. CONCLUSION: There was a clear tradeoff between adhesive strength and toxicity. The maximum failure strength in shear or tension of the three surgical adhesives ranked from strongest to the weakest was: 1) Glubran 2, 2) BioGlue, and 3) Tisseel. Tisseel was found to be the least toxic of the three adhesives, whereas Glubran 2 was the most toxic. LEVEL OF EVIDENCE: NA Laryngoscope, 129:2139-2146, 2019.

来源:Laryngoscope,查看原文。版权归原作者所有。