水肺潜水时反复高压暴露对人工耳蜗的影响
本文通过回顾性病例系列和文献综述,描述爱尔兰国家听觉植入与研究中心3例共5个耳蜗植入且潜水经验丰富的患者,其中2例需再植入(一例因非听觉刺激,一例因反复水肺潜水后电极经鼓膜挤出),第三例80次潜水后无并发症;作者建议常规水肺潜水的人工耳蜗患者谨慎并严格遵循安全限制。
要点
- 人工耳蜗患者水肺潜水可能出现植入体相关并发症,装置故障为理论风险。
- 鼓阶开窗可致外淋巴外渗、内耳气压伤、减压病及电极周围气泡形成。
- 作者建议常规水肺潜水的人工耳蜗患者谨慎并严格遵循推荐的安全限制。
核心结论
- 人工耳蜗患者水肺潜水可能出现植入体相关并发症,装置故障为理论风险;作者建议常规水肺潜水的人工耳蜗患者谨慎并严格遵循推荐的安全限制。
研究设计与样本
- 回顾性病例系列和文献综述。
- 回顾爱尔兰国家听觉植入与研究中心已知的水肺潜水患者,并通过PubMed和Google Scholar检索文献。
主要发现
- 3例水肺潜水的人工耳蜗患者中,2例需再植入:第一例因非听觉刺激,第二例因反复水肺潜水后电极经鼓膜挤出。
- 第三例患者在80次潜水后仍无并发症。
- 鼓阶开窗可导致外淋巴外渗、内耳气压伤、减压病及电极周围气泡形成。
- 耳聋、前庭病变伴突发外淋巴漏及本体感觉丧失可导致定向障碍和蓝顶综合征。
临床怎么用
- 作者基于本中心水肺潜水人工耳蜗患者经验及文献报道,建议常规水肺潜水的人工耳蜗患者谨慎并严格遵循推荐的安全限制。
局限性
- 证据水平为4。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES/HYPOTHESIS: Complications during or after cochlear implantation are relatively rare. They occur more frequently in patients who partake in activities that can potentially lead to local trauma. No formal recommendations exist for participation in self-contained underwater breathing apparatus (SCUBA) activities. We describe three patients with a combined five cochlear implants and extensive diving experience, the largest case series to date, and highlight some of the difficulties faced. We also review the literature on previously described SCUBA-diving patients with cochlear implants. STUDY DESIGN: Retrospective case series and literature review. METHODS: A review of the known SCUBA divers in the National Hearing Implant and Research Centre in Ireland was conducted, and a review of the literature was carried out using PubMed and Google Scholar. RESULTS: Of the three SCUBA divers with cochlear implants, two required reimplantation, the first due to nonauditory stimulation, and the second due to extrusion of the electrode through the tympanic membrane following repetitive SCUBA dives. The third patient remains without complications after 80 dives. CONCLUSIONS: Patients with cochlear implants can have complications relating to the implant itself, with device failure a theoretical risk. The cochleostomy can lead to perilymphatic extravasation, as well as inner ear barotrauma, decompression sickness, and formation of air bubbles along the electrode. A combination of deafness, vestibulopathy with abrupt perilymph leak, and loss of proprioception can lead to disorientation and blue dome syndrome. Based on our experience with cochlear implants in SCUBA divers, along with those reported in the literature, we recommend caution in patients with cochlear implants who SCUBA dive regularly and strict adherence to the recommended safety limits. LEVEL OF EVIDENCE: 4 129:2760-2764, 2019.
引用
- Hintze JM, Geyer L, Fitzgerald CW, et al. The impact of repetitive hyperbaric exposure during SCUBA diving on cochlear implants. Laryngoscope. 2019. PMID: 30810235. doi:10.1002/lary.27880