3D打印模型中活塞假体放置时施加的力:徒手与机器人辅助技术的比较
前瞻性可行性研究在3D打印中耳模型上比较徒手与机器人耳鼻喉显微手术系统(REMS)放置和夹持活塞假体时对砧骨施加的最大力,6名参与者各完成3次徒手和3次REMS辅助操作,结果显示REMS辅助夹持时最大力显著低于徒手,放置时无显著差异。
要点
- REMS辅助夹持活塞假体时对砧骨的最大力显著低于徒手技术(P = .049)。
- 徒手与REMS辅助放置假体时最大力无显著差异(P = .404)。
- REMS辅助放置和夹持活塞假体在3D打印模型中可行。
核心结论
- 初步数据显示REMS辅助放置和夹持镫骨活塞假体可行,且与徒手相比REMS夹持时对砧骨施加的最大力显著降低。
研究设计与样本
- 前瞻性可行性研究,在机器人实验室进行,6名参与者(1名神经耳科专科培训医师、2名神经耳科研究员、3名耳鼻咽喉头颈外科住院医师)在3D打印中耳模型上各完成3次徒手和3次REMS辅助的活塞假体放置与夹持操作。
主要发现
- 假体放置时平均最大力:徒手组126.4 ± 73.6 mN vs REMS组105.0 ± 69.4 mN(P = .404)。
- 假体夹持时平均最大力:徒手组469.3 ± 225.2 mN vs REMS组272.7 ± 97.4 mN(P = .049)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: To describe a 3D-printed middle ear model that quantifies the force applied to the modeled incus. To compare the forces applied during placement and crimping of a stapes prosthesis between the Robotic ENT Microsurgery System ( REMS) and the freehand technique in this model. STUDY DESIGN: Prospective feasibility study. SETTING: Robotics laboratory. SUBJECTS AND METHODS: A middle ear model was designed and 3D printed to facilitate placement and crimping of a piston prosthesis. The modeled incus was mounted to a 6-degree of freedom force sensor to measure forces/torques applied on the incus. Six participants-1 fellowship-trained neurotologist, 2 neurotology fellows, and 3 otolaryngology-head and neck surgery residents-placed and crimped a piston prosthesis in this model, 3 times freehand and 3 times REMS assisted. Maximum force applied to the incus was then calculated for prosthesis placement and crimping from force/torque sensor readings for each trial. Robotic and freehand outcomes were compared with a linear regression model. RESULTS: Mean maximum magnitude of force during prosthesis placement was 126.4 ± 73.6 mN and 105.0 ± 69.4 mN for the freehand and robotic techniques, respectively ( P = .404). For prosthesis crimping, the mean maximum magnitude of force was 469.3 ± 225.2 mN for the freehand technique and 272.7 ± 97.4 mN for the robotic technique ( P = .049). CONCLUSIONS: Preliminary data demonstrate that REMS-assisted stapes prosthesis placement and crimping are feasible with a significant reduction in maximum force applied to the incus during crimping with the REMS in comparison with freehand.
引用
- Razavi CR, Wilkening PR, Yin R, et al. Applied Force during Piston Prosthesis Placement in a 3D-Printed Model: Freehand vs Robot-Assisted Techniques. Otolaryngol Head Neck Surg. 2018. PMID: 30509152. doi:10.1177/0194599818815144
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。