主动中耳植入体听力康复的长期结局
结果显示装置平均使用6.7年,77.12%患者观察期末仍在使用,总体并发症率16.1%,装置故障率3.4%,各型听力损失患者均有显著听力增益。
要点
- 主动中耳植入体长期使用可靠,多数患者观察期末仍在使用。
- 总体并发症率16.1%,装置故障率3.4%,可接受。
- 各型听力损失患者术后听力学获益显著。
核心结论
- 长期随访显示主动中耳植入体在单中心应用中具有可靠性,总体并发症率与装置故障率可接受。
- 各型听力损失患者的听力学获益均满意,多数患者观察期末仍在使用植入体。
研究设计与样本
- 回顾性病历分析,纳入1998年9月1日至2015年7月31日单中心植入主动中耳植入体的所有患者。
主要发现
- 总体并发症率16.1%;2004至2006年间植入装置的翻修率和取出率较高。
- 装置故障率3.4%;听力学评估显示两类听力损失患者均有显著听力增益。
关键数值与推荐
- 功能增益评估覆盖250至4000 Hz各频率。
- 言语识别采用65 dB下弗莱堡单音节词测试。
临床怎么用
- 主动中耳植入体可作为感音神经性、混合性及传导性听力损失的长期听力康复选择。
- 需关注并发症与装置故障,尤其在使用替代耦合方式时应加强随访。
局限性
- 本研究为单中心回顾性病历分析,证据等级为4级。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To assess the audiological and long-term medical and technical follow-up outcomes of an active middle ear implant. METHODS: This was a retrospective medical chart analysis of all patients provided with an active middle ear implant in a tertiary academic medical referral center between September 1, 1998, and July 31, 2015. Main outcome measures were medical and technical complications, revisions, reimplantations, explantations, coupling approaches, mean time of use, pre- and postoperative hearing thresholds, functional hearing gain across frequencies (250-4,000 Hz), and Freiburg monosyllablic word test at 65 dB. RESULTS: One hundred and three patients were identified. Fifteen were implanted bilaterally (n = 118 Vibrant Soundbridge devices [MED-EL, Innsbruck, Austria]). Seventy-seven devices were implanted for sensorineural and 41 for mixed and conductive hearing loss. Patients used the implant for 6.7 years (range 0.7 months-17.9 years) on average. Ninety-one patients (77.12%) were using the device at the end of the observation period. An overall complication rate of 16.1% was observed. The revision and explantation rates were higher for devices implanted between 2004 and 2006. The device failure rate was 3.4%. Audiological evaluation showed significant hearing gains for both hearing loss patient groups. CONCLUSION: This long-term follow-up reveals the reliability of the active middle ear implant in a single center. Overall complication rate and device failure rate are acceptable. The complication rate was higher during implementation of alternative coupling approaches. The audiological benefit was satisfactory in patients with all hearing loss types. The majority of implanted patients used the implant at the end of the observation period. LEVEL OF EVIDENCE: 4 Laryngoscope, 129:477-481, 2019.
引用
- Brkic FF, Riss D, Auinger A, et al. Long-Term Outcome of Hearing Rehabilitation With An Active Middle Ear Implant. Laryngoscope. 2018. PMID: 30284273. doi:10.1002/lary.27513
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。