探测音频率对儿童人工耳蜗植入者同侧与对侧电诱发镫骨肌反射测量的影响

Effect of Probe-Tone Frequency on Ipsilateral and Contralateral Electrical Stapedius Reflex Measurement in Children With Cochlear Implants.
PubMed 文献 · Ear Hear · 2019-01-01

研究纳入19例单侧Advanced Bionics人工耳蜗植入儿童,比较226、678、1000 Hz探测音在同侧与对侧测量电诱发镫骨肌反射阈值(eSRT)的成功率。结果显示同侧eSRT成功率随探测音频率升高而显著增加,对侧无显著差异;平均eSRT水平在各条件下无显著差异。

要点

  • 同侧eSRT成功率在1000 Hz探测音下高于226 Hz(37%对16%)。
  • 对侧eSRT成功率在各探测音频率间无显著差异。
  • 平均eSRT水平不随探测音频率或测量耳别变化。

核心结论

  • 使用较高探测音频率(如1000 Hz)可提高儿童人工耳蜗植入者同侧eSRT测量成功率。

研究设计与样本

  • 研究纳入19例单侧Advanced Bionics人工耳蜗植入儿童(平均年龄8.6岁,SD=2.29),使用整合中耳分析仪在226、678、1000 Hz探测音下测量同侧与对侧eSRT。

主要发现

  • 同侧eSRT在226、678、1000 Hz探测音下分别成功测量3耳(16%)、4耳(21%)、7耳(37%)。
  • 对侧eSRT在226、678、1000 Hz探测音下分别成功测量11耳(58%)、13耳(68%)、13耳(68%)。
  • 同侧eSRT成功测量率随探测音频率变化有显著差异,226 Hz与1000 Hz间差异最大。
  • 对侧eSRT成功测量率随探测音频率变化无显著差异。
  • 平均eSRT水平在同侧与对侧测量间及不同探测音频率间均无显著差异。

临床怎么用

  • 对儿童人工耳蜗植入者进行同侧eSRT测量时,可优先尝试1000 Hz探测音以提高成功率。

局限性

  • 研究样本量较小(19例),且仅纳入单侧Advanced Bionics人工耳蜗植入儿童。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The upper loudness limit of electrical stimulation in cochlear implant patients is sometimes set using electrically elicited stapedius reflex thresholds (eSRTs), especially in children for whom reporting skills may be limited. In unilateral cochlear implant patients, eSRT levels are measured typically in the contralateral unimplanted ear because the ability to measure eSRTs in the implanted ear is likely to be limited due to the cochlear implant surgery and consequential changes in middle ear dynamics. This practice is particularly limiting in the case of fitting bilaterally implanted pediatric cases because there is no unimplanted ear option to choose for eSRT measurement. The goal of this study was to identify an improved measurement protocol to increase the success of eSRT measurement in ipsilateral or contralateral or both implanted ears of pediatric cochlear implant recipients. This work hypothesizes that use of a higher probe frequency (e.g., 1000 Hz compared with the 226 Hz standard), which is closer to the mechanical middle ear resonant frequency, may be more effective in measuring middle ear muscle contraction in either ear. DESIGN: In the present study, eSRTs were measured using multiple probe frequencies (226, 678, and 1000 Hz) in the ipsilateral and contralateral ears of 19 children with unilateral Advanced Bionics (AB) cochlear implants (mean age = 8.6 years, SD = 2.29). An integrated middle ear analyzer designed by AB was used to elicit and detect stapedius reflexes and assign eSRT levels. In the integrated middle ear analyzer system, an Interacoustics Titan middle ear analyzer was used to perform middle ear measurements in synchrony with research software running on an AB Neptune speech processor, which controlled the delivery of electrical pulse trains at varying levels to the test subject. Changes in middle ear acoustic admittance following an electrical pulse train stimulus indicated the occurrence of an electrically elicited stapedius reflex. RESULTS: Of the 19 ears tested, ipsilateral eSRTs were successfully measured in 3 (16%), 4 (21%), and 7 (37%) ears using probe tones of 226, 678, and 1000 Hz, respectively. Contralateral eSRT levels were measured in 11 (58%), 13 (68%), and 13 (68%) ears using the three different probe frequencies, respectively. A significant difference was found in the incidence of successful eSRT measurement as a function of probe frequency in the ipsilateral ears with the greatest pair-wise difference between the 226 and 1000 Hz probe. A significant increase in contralateral eSRT measurement success as a function of probe frequency was not found. These findings are consistent with the idea that changes in middle ear mechanics, secondary to cochlear implant surgery, may interfere with the detection of stapedius muscle contraction in the ipsilateral middle ear. The best logistic, mixed-effects model of the occurrence of successful eSRT measures included ear of measurement and probe frequency as significant fixed effects. No significant differences in average eSRT levels were observed across ipsilateral and contralateral measurements or as a function of probe frequency. CONCLUSION: Typically, measurement of stapedius reflexes is less successful in the implanted ears of cochlear implant recipients compared with measurements in the contralateral, unimplanted ear. The ability to measure eSRT levels ipsilaterally can be improved by using a higher probe frequency.