主动中耳植入物的气导与骨导反馈来源研究
该研究在4例新鲜冷冻半侧人头标本上制备乳突与上鼓室切开,将Carina主动中耳植入物致动器以两种方式耦合于砧骨,用纯音500–6000 Hz驱动,测量耳道声压与乳突骨/皮肤振动。结果显示耳道声压达60–89 dB SPL,乳突骨振动峰值高于皮肤,提示气导与骨导信号均足以被植入麦克风检测并可能再放大。
要点
- 主动中耳植入物可产生气导与骨导两种反馈信号
- 耳道声压达60–89 dB SPL,峰值在2 kHz以下
- 乳突骨振动大于皮肤振动,骨导通路不可忽视
核心结论
- 主动中耳植入物产生的气导与骨导信号强度均足以被植入装置麦克风检测,可能导致再放大。
研究设计与样本
- 研究在4例新鲜冷冻半侧人头标本上制备乳突与上鼓室切开,将Carina主动中耳植入物致动器以立体定向支架和乳突皮质骨固定支架两种方式耦合于砧骨,用500–6000 Hz纯音以1/4倍频程步进驱动,测量耳道声压与骨导振动。
主要发现
- 耳道声压为60–89 dB SPL,峰值在2 kHz以下频率范围。
- 乳突骨激光多普勒振动峰值:固定支架组0.32–0.79 mm/s,立体定向悬吊组0.21–0.36 mm/s;皮肤峰值:固定支架组0.05–0.15 mm/s,立体定向悬吊组0.03–0.13 mm/s。
临床怎么用
- 理解气导与骨导反馈的相对贡献,有助于开发针对性抑制算法及强调声隔离的手术技术。
局限性
- 研究基于4例新鲜冷冻人头标本,为离体实验,结果向活体临床转化需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Active middle ear implants (AMEI) have been used to treat hearing loss in patients for whom conventional hearing aids are unsuccessful for varied biologic or personal reasons. Several studies have discussed feedback as a potential complication of AMEI usage, though the feedback pathway is not well understood. While reverse propagation of an acoustic signal through the ossicular chain and tympanic membrane constitutes an air-conducted source of feedback, the implanted nature of the device microphone near the mastoid cortex suggests that bone conduction pathways may potentially be another significant factor. This study examines the relative contributions of potential sources of feedback during stimulation with an AMEI. DESIGN: Four fresh-frozen, hemi-sectioned, human cadaver specimens were prepared with a mastoid antrostomy and atticotomy to visualize the posterior incus body. A Carina active middle ear implant actuator (Cochlear Ltd., Boulder, CO) was coupled to the incus by two means: (1) a stereotactic arm mounted independently of the specimen and (2) a fixation bracket anchored directly to the mastoid cortical bone. The actuator was driven with pure-tone frequencies in 1/4 octave steps from 500 to 6000 Hz. Acoustic sound intensity in the ear canal was measured with a probe tube microphone (Bruel & Kjær, Nærum, Denmark). Bone-conducted vibration was quantified with a single-axis laser Doppler vibrometer (Polytec Inc., Irvine, CA) from both a piece of reflective tape placed on the skin overlying the mastoid and a bone-anchored titanium screw and pedestal (Cochlear Ltd., Centennial, CO) implanted in the cortical mastoid bone. RESULTS: Microphone measurements revealed ear-canal pressures of 60-89 dB SPL, peaking in the frequency range below 2 kHz. Peak LDV measurements were greatest on the mastoid bone (0.32-0.79 mm/s with mounting bracket and 0.21-0.36 mm/s with the stereotactic suspension); peak measurements on the skin ranged from 0.05 to 0.15 mm/s with the bracket and 0.03 to 0.13 mm/s with stereotactic suspension. CONCLUSION: AMEI produce both air- and bone-conducted signals of adequate strength to be detected by the implanted device microphone, potentially resulting in reamplification. Understanding the relative contribution of these sources may play an important role in the development of targeted mitigation algorithms, as well as surgical techniques emphasizing acoustic isolation.
引用
- Banakis Hartl RM, Easter JR, Alhussaini MA, et al. Air- and Bone-Conducted Sources of Feedback With an Active Middle Ear Implant. Ear Hear. 2019. PMID: 30199397. doi:10.1097/AUD.0000000000000655
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。