文献速递耳科与耳神经

优化音高与语言策略在人工耳蜗使用者中的评估

Evaluation of the Optimized Pitch and Language Strategy in Cochlear Implant Recipients.
PubMed 文献 · Ear Hear · 2019-01-01

本研究在20名成人人工耳蜗使用者中比较OPAL策略与临床ACE策略,18套数据纳入分析。OPAL在言语语调感知上较ACE平均提高8.5个百分点,安静与噪声下言语感知在适应期后相当,验配流程无需额外步骤。

要点

  • OPAL较ACE在语调感知平均提高8.5个百分点
  • 噪声下句识别需约4周适应期后才与ACE相当
  • OPAL可使用与ACE相同的验配流程,默认F0恢复增益与自定义增益效果相当

核心结论

  • CP900 OPAL实现较ACE显著改善言语语调感知,适应后安静与噪声下言语感知及主观言语清晰度评分相当,验配流程与ACE相同。

研究设计与样本

  • 重复测量单受试者设计并结合居家使用体验,纳入20名有经验的成人人工耳蜗使用者,2名退出后分析18套数据。

主要发现

  • OPAL较ACE在语调测试中组平均获益8.5个百分点。
  • OPAL在急性期及仅2周居家体验后噪声下句识别显著较差,4周居家体验后安静下词识别与噪声下句识别与ACE相当。
  • 现场受试者接受度良好,两策略言语清晰度评分相当。
  • OPAL未增加验配步骤,默认F0恢复增益与自定义增益结果相当。

关键数值与推荐

  • OPAL刺激速率为1200 pulses per second/channel,ACE为900 pulses per second/channel。
  • 语调测试组平均获益为8.5% points(OPAL对比ACE)。
  • 适应期观察点为急性期、2周与4周居家体验。

临床怎么用

  • OPAL可在常规ACE验配流程下编程,采用默认F0恢复增益即可,但需告知患者噪声下言语感知存在约4周适应期。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The Optimized Pitch and Language (OPAL) strategy enhances pitch perception through coding of fundamental frequency (F0) amplitude modulation information in the stimulus envelope delivered to a cochlear implant. Previous research using a prototype of the strategy demonstrated significant benefits in musical pitch and lexical tone discrimination tasks with no degradation in speech recognition when compared with the clinical Advanced Combination Encoder (ACE) strategy in a small group of subjects. Based on those studies, a modified version of the strategy was implemented in the commercial Nucleus CP900 series processor. The aims of the present study were to establish whether the CP900 OPAL implementation continued to provide improved F0 pitch perception in a speech intonation task with no degradation to speech perception in quiet and noise, when compared with the clinical ACE strategy in a larger cohort of subjects. Further aims were to evaluate fitting procedures and subject acclimatization to the strategy after take-home experience. DESIGN: Twenty experienced adult cochlear implant recipients were enrolled in the study. Two subjects withdrew during the study leaving 18 sets of data for analysis. A repeated-measures single-subject design with take-home experience was used to test for improved speech intonation perception using OPAL compared with ACE and for comparable performance between strategies for open-set word recognition in quiet at two presentation levels, sentence recognition in adaptive 4-talker babble noise, and speech intelligibility ratings. The stimulation rate employed for OPAL was 1200 pulses per second/channel which was higher than the default clinical rate of 900 pulses per second/channel used for ACE by all subjects in the present study. Two variations of the OPAL "F0 restore gain" (the gain applied to restore the loudness of modulated channels) were investigated: "custom" measured per subject and "default" which was the average of all subject custom gains. RESULTS: A significant group mean benefit on the intonation test of 8.5% points was shown for OPAL compared with ACE. There was a significant period of adaptation to OPAL with significantly poorer sentence in noise scores acutely and after only 2 weeks of take-home experience. After 4 weeks of take-home experience, comparable word perception in quiet and sentence perception in noise for OPAL were obtained. Furthermore, there was good subject acceptability in the field with comparable speech intelligibility ratings between strategies. Results of the fitting procedure showed that OPAL did not require any additional steps compared with fitting of ACE. A default F0 restore gain provided comparable outcomes to a custom gain setting. CONCLUSIONS: The CP900 OPAL implementation provided a significant benefit to perception of speech intonation when compared with ACE. Comparable speech perception (in quiet and noise) and subjective ratings of speech intelligibility between strategies were also achieved after a period of acclimatization. These outcomes are consistent with results of earlier studies using prototype versions of the strategy and reaffirm its potential for improvement of F0 pitch perception in speech while preserving coding of segmental speech information. Furthermore, the OPAL strategy can be programmed into subject's processors using the same fitting procedures used for ACE thereby simplifying its adoption in clinical settings.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Vandali A, Dawson P, Au A, et al. Evaluation of the Optimized Pitch and Language Strategy in Cochlear Implant Recipients. Ear Hear. 2019. PMID: 30067558. doi:10.1097/AUD.0000000000000627

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。