文献速递耳科与耳神经

动态电流聚焦:人工耳蜗响度编码新方法

Dynamic Current Focusing: A Novel Approach to Loudness Coding in Cochlear Implants.
PubMed 文献 · Ear Hear · 2019-01-01

本研究提出一种动态电流聚焦(DCF)策略,在近阈值处使用三极刺激并逐渐增宽兴奋范围以增加响度而不增加总电流。11名成人耳蜗植入者参与,DCF在45 dB下显著改善频谱纹波分辨,但言语识别与时间调制检测无差异,多数无主观偏好。

要点

  • DCF策略在低响度水平改善频谱分辨率。
  • DCF与单极策略在言语识别和时域处理上等效。
  • DCF增加耗电,电池寿命降至1.5–4小时。

核心结论

  • DCF策略在较低响度水平提供更好的频谱分辨率,但言语测试表现与单极策略相当。

研究设计与样本

  • 研究纳入11名有至少9个月HiRes90K使用经验的成人语后聋耳蜗植入者,比较DCF研究处理器与临床Harmony处理器。

主要发现

  • 在45 dB下,DCF的平均频谱纹波得分从临床程序的2.4 ripples/octave显著提高至3.74 ripples/octave(p=0.016)。
  • 11名参与者中有8名在65 dB下频谱分辨率改善,但DCF与单极在更高呈现水平无显著差异。
  • 言语噪声下识别和时间调制检测阈值在两种策略间相等。
  • 主观上2名偏好DCF,2名偏好自身程序,多数无偏好;电池寿命减少至1.5–4小时。

关键数值与推荐

  • 阈值确定时σ=0.8,即80%电流返回侧翼电极,20%返回耳外接地电极。
  • 通过降低σ而非增加脉冲幅度来确定最舒适响度。

临床怎么用

  • DCF可能作为改善低响度下频谱分辨率的选项,但需考虑耗电增加且言语获益未证实。

局限性

  • 适应期仅2–3小时,需更长适应期研究长期效果。
  • 样本量小(11人),且频谱纹波改善是否转移至言语得分未知。
  • 需进一步研究 fitting 规则和降低功耗以适合常规临床。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: In an attempt to improve spectral resolution and speech intelligibility, several current focusing methods have been proposed to increase spatial selectivity by decreasing intracochlear current spread. For example, tripolar stimulation administers current to a central electrode and uses the two flanking electrodes as the return pathway, creating a narrower intracochlear electrical field and hence increases spectral resolution when compared with monopolar (MP) stimulation. However, more current is required, and in some patients, specifically the ones with high electrode impedances, full loudness growth cannot be supported because of compliance limits. The present study describes and analyses a new loudness encoding approach that uses tripolar stimulation near threshold and gradually broadens the excitation (by decreasing compensation coefficient σ) to increase loudness without the need to increase overall current. It is hypothesized that this dynamic current focusing (DCF) strategy increases spatial selectivity, especially at lower loudness levels, while maintaining maximum selectivity at higher loudness levels, without reaching compliance limits. DESIGN: Eleven adult cochlear implant recipients with postlingual hearing loss, with at least 9 months of experience with their HiRes90K implant, were selected to participate in this study. Baseline performance regarding speech intelligibility in noise (Dutch matrix sentence test), spectral ripple discrimination at 45 and 65 dB, and temporal modulation detection thresholds were assessed using their own clinical program, fitted on a Harmony processor. Subsequently, the DCF strategy was fitted on a research Harmony processor. Threshold levels were determined with σ = 0.8, which means 80% of current is returned to the flanking electrodes and the remaining 20% to the extracochlear ground electrode. Instead of increasing overall pulse magnitude, σ was decreased to determine most comfortable loudness. After 2 to 3 hr of adaptation to the research strategy, the same psychophysical measures were taken. RESULTS: At 45 dB, average spectral ripple scores improved significantly from 2.4 ripples per octave with their clinical program to 3.74 ripples per octave with the DCF strategy (p = 0.016). Eight out of 11 participants had an improved spectral resolution at 65 dB. Nevertheless, no significant difference between DCF and MP was observed at higher presentation levels. Both speech-in-noise and temporal modulation detection thresholds were equal for MP and DCF strategies. Subjectively, 2 participants preferred the DCF strategy over their own clinical program, 2 preferred their own strategy, while the majority of the participants had no preference. Battery life was decreased and ranged from 1.5 to 4 hr. CONCLUSIONS: The DCF strategy gives better spectral resolution, at lower loudness levels, but equal performance on speech tests. These outcomes warrant for a longer adaptation period to study long-term outcomes and evaluate if the outcomes in the ripple tests transfer to the speech scores. Further research, for example, with respect to fitting rules and reduction of power consumption, is necessary to make the DCF strategy suitable for routine clinical application.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. de Jong MAM, Briaire JJ, Frijns JHM, et al. Dynamic Current Focusing: A Novel Approach to Loudness Coding in Cochlear Implants. Ear Hear. 2019. PMID: 29742542. doi:10.1097/AUD.0000000000000593

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。