舌运动变异性与上气道刺激电极配置变化及对治疗结果的影响
这项两中心前瞻性 consecutive 试验分析了35例使用上气道刺激(UAS)至少12个月的患者,评估不同电极配置下的舌运动模式及其与治疗反应的相关性。结果显示UAS显著降低OSA严重程度(P < .001),58.8%患者在不同电极配置下出现舌运动模式变化,73.5%在右侧与双侧前伸间交替,且切换电极设置时需不同电压达到功能刺激水平。
要点
- UAS治疗可显著降低OSA严重程度(P < .001)。
- 不同电极配置下58.8%患者舌运动模式发生变化。
- 73.5%患者在右侧与双侧前伸间交替,被视为选择性UAS反应的最佳表型。
核心结论
- UAS对特定OSA患者高度有效,但需关注电极配置变化引起的舌运动变化。
研究设计与样本
- 两中心、前瞻性 consecutive 试验,纳入35例设备使用至少12个月的患者。
主要发现
- UAS治疗显著降低OSA严重程度(P < .001)。
- 不同电极配置下舌运动模式变化常见(58.8%)。
- 73.5%患者在右侧与双侧前伸间交替,属选择性UAS反应的最佳表型。
- 切换电极设置时需不同电压达到功能刺激水平。
关键数值与推荐
- 纳入患者呼吸暂停低通气指数在15至65次/小时之间且体重指数较高。
- 证据等级为4。
临床怎么用
- 对电极配置变化后舌运动发生偏移的患者应予以关注。
- 观察到舌运动偏移时,应重新评估术中袖带放置。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES/HYPOTHESIS: Upper airway stimulation (UAS) is an effective treatment for obstructive sleep apnea (OSA). Previous data have demonstrated a correlation between the phenotype of tongue motion and therapy response. Closed loop hypoglossal nerve stimulation implant offers five different electrode configuration settings which may result in different tongue motion. STUDY DESIGN: Two-center, prospective consecutive trial in a university hospital setting. METHODS: Clinical outcomes of 35 patients were analyzed after at least 12 months of device use. Tongue motion was assessed at various electrode configuration settings. Correlation between the tongue motion and treatment response was evaluated. RESULTS: OSA severity was significantly reduced with the use of UAS therapy (P < .001). Changes in tongue motion patterns were frequently observed (58.8%) with different electrode configuration settings. Most of the patients alternated between right and bilateral protrusion (73.5%), which are considered to be the optimal phenotypes for selective UAS responses. Different voltage settings were required to achieve functional stimulation levels when changing between the electrode settings. CONCLUSIONS: UAS is highly effective for OSA treatment in selected patients with an apnea-hypopnea index between 15 and 65 events per hour and higher body mass index. Attention should be given to patients with shifting tongue movement in response to change of electrode configuration. The intraoperative cuff placement should be reassessed when tongue movement shifting is observed. LEVEL OF EVIDENCE: 4 Laryngoscope, 1970-1976, 2018.
引用
- Steffen A, Kilic A, König IR, et al. Tongue motion variability with changes of upper airway stimulation electrode configuration and effects on treatment outcomes. Laryngoscope. 2017. PMID: 29280488. doi:10.1002/lary.27064
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。