总和电位是耳蜗电图电极位置可靠标志:人工耳蜗作为诊疗探针

The Summating Potential Is a Reliable Marker of Electrode Position in Electrocochleography: Cochlear Implant as a Theragnostic Probe.
PubMed 文献 · Ear Hear · 2018-01-01

该研究在10只听力正常的Dunkin Hartley豚鼠中植入定制六触点人工耳蜗,术中经各触点记录耳蜗电图并分离各成分,结合5个耳蜗的组织学重建验证。结果显示总和电位(SP)可标记记录电极的频率位置,精度约±1/4倍频程,并经组织学验证;耳蜗微音电位可靠性较低,复合动作电位不适合定位但对耳蜗健康敏感。

要点

  • 术中经人工耳蜗触点记录的总和电位可标记电极的耳蜗频率位置,精度约±1/4倍频程。
  • 耳蜗微音电位虽与电极位置相关,但作为位置标志可靠性较低。
  • 复合动作电位不能提供电极位置信息,但对插入创伤等耳蜗健康状态敏感。

核心结论

  • 经人工耳蜗触点术中记录的总和电位可用于判断电极在耳蜗内的频率位置。
  • 若该技术在人类中同样有效,或可在损伤功能性耳蜗区域前停止植入,从而减少植入创伤、提高听力保留。

研究设计与样本

  • 在10只听力正常(听性脑干反应阈值正常)的Dunkin Hartley豚鼠中行耳蜗内耳蜗电图测量。
  • 全部动物植入定制六触点人工耳蜗(深4至5 mm),以单极配置(耳后参考电极)同步记录六个触点信号。
  • 离线滤波分离复合动作电位、耳蜗微音电位和总和电位三种主要成分;另取5个耳蜗行组织学处理以评估触点空间位置。
  • 耳蜗电图数据与组织学电极位置重建均以Greenwood函数拟合,以验证推算的耳蜗频率位置可靠性。

主要发现

  • 总和电位可作为记录电极频率位置的合适标志,在功能正常的耳蜗中精度为±1/4倍频程,并经组织学验证。
  • 耳蜗微音电位显示对电极位置有依赖性,但作为位置标志的可靠性较低。
  • 复合动作电位不适合提供人工耳蜗位置信息,但对「耳蜗健康」(如插入创伤)敏感。

临床怎么用

  • 术中可利用人工耳蜗触点记录总和电位,监测电极的耳蜗频率位置,辅助判断是否应停止继续植入。
  • 复合动作电位信号可用于提示插入创伤等耳蜗健康问题。
  • 若能证实该技术在人类中同样有效,或有助于减少植入创伤、提高听力保留。
  • 具备无滤波限制记录生物信号能力的诊断硬件与软件,可能成为人工耳蜗制造商产品组合的有价值补充。

局限性

  • 研究为动物实验(10只豚鼠),结论外推至人类需进一步验证。
  • 原文以「若该技术在人类中同样有效」表述其临床价值,尚未在人类中证实。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: For the increasing number of cochlear implantations in subjects with residual hearing, hearing preservation, and thus the prevention of implantation trauma, is crucial. A method for monitoring the intracochlear position of a cochlear implant (CI) and early indication of imminent cochlear trauma would help to assist the surgeon to achieve this goal. The aim of this study was to evaluate the reliability of the different electric components recorded by an intracochlear electrocochleography (ECochG) as markers for the cochleotopic position of a CI. The measurements were made directly from the CI, combining intrasurgical diagnostics with the therapeutical use of the CI, thus, turning the CI into a "theragnostic probe." DESIGN: Intracochlear ECochGs were measured in 10 Dunkin Hartley guinea pigs of either sex, with normal auditory brainstem response thresholds. All subjects were fully implanted (4 to 5 mm) with a custom six contact CI. The ECochG was recorded simultaneously from all six contacts with monopolar configuration (retroauricular reference electrode). The gross ECochG signal was filtered off-line to separate three of its main components: compound action potential, cochlear microphonic, and summating potential (SP). Additionally, five cochleae were harvested and histologically processed to access the spatial position of the CI contacts. Both ECochG data and histological reconstructions of the electrode position were fitted with the Greenwood function to verify the reliability of the deduced cochleotopic position of the CI. RESULTS: SPs could be used as suitable markers for the frequency position of the recording electrode with an accuracy of ±1/4 octave in the functioning cochlea, verified by histology. Cochlear microphonics showed a dependency on electrode position but were less reliable as positional markers. Compound action potentials were not suitable for CI position information but were sensitive to "cochlear health" (e.g., insertion trauma). CONCLUSIONS: SPs directly recorded from the contacts of a CI during surgery can be used to access the intracochlear frequency position of the CI. Using SP monitoring, implantation may be stopped before penetrating functioning cochlear regions. If the technique was similarly effective in humans, it could prevent implantation trauma and increase hearing preservation during CI surgery. Diagnostic hardware and software for recording biological signals with a CI without filter limitations might be a valuable add-on to the portfolios of CI manufacturers.