文献速递耳科与耳神经

振幅增长函数推导eCAP阈值的精确性研究

The Precision of eCAP Thresholds Derived From Amplitude Growth Functions.
PubMed 文献 · Ear Hear · 2018-01-01

该回顾性研究基于111例植入HiRes90K人工耳蜗患者的826条术中振幅增长函数,用线性外推法和最后可见eCAP法确定eCAP阈值并计算阈值置信区间(TCI)评估精确性。结果显示TCI常大于预期,eCAP阈值可能不足以精确预测行为T轮廓。

要点

  • 线性外推法的TCI大于最后可见法,且阈值系统性偏小。
  • 组分析中eCAP与行为阈值相关系数r=0.44(线性外推)和r=0.54(最后可见)。
  • 个体内相关系数约-1至+1,提示个体水平预测不稳定。

核心结论

  • eCAP阈值存在可用TCI估计的不确定性,TCI常大于仅看阈值时的预期。
  • eCAP阈值可能不够精确以预测行为T轮廓,未来研究应伴随精确性指标。

研究设计与样本

  • 回顾性研究,纳入111例植入HiRes90K人工耳蜗的患者,共826条术中测量的振幅增长函数。
  • 每条AGF用线性外推法和最后可见eCAP法确定eCAP阈值,并计算阈值置信区间(TCI)作为精确性指标。

主要发现

  • 线性外推法的TCI大于最后可见法,且线性外推法估计的eCAP阈值系统性小于最后可见法。
  • 组分析中eCAP与行为阈值的相关系数:线性外推法r=0.44,最后可见法r=0.54。
  • 个体内相关系数在两种方法下均约从-1至+1变化。
  • 跨受试者分析显示行为阈值落在eCAP阈值轮廓的TCI内。

关键数值与推荐

  • TCI越大表示精确性或可靠性越低。
  • 线性外推法中TCI最小的阈值与最后可见法阈值相关性最好。

临床怎么用

  • 临床使用eCAP阈值时应考虑其TCI,不宜仅凭阈值数值判断精确性。
  • eCAP阈值可能不足以精确预测行为T轮廓,需结合行为测听结果。

局限性

  • 研究为回顾性设计,仅纳入HiRes90K人工耳蜗患者。
  • 未报告TCI的具体数值范围或影响因素定量分析。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: An amplitude growth function (AGF) shows the amplitude of an electrically evoked compound action potential (eCAP) as a function of the stimulation current. AGFs can be used to derive the eCAP threshold, which represents the minimum amount of current needed to elicit a measurable eCAP. eCAP thresholds have been widely used clinically to, for example, assist with sound processor programming. However, no eCAP precision has been included to date. The aim of this study was to investigate the precision of eCAP thresholds and determine whether they are precise enough for clinical use. DESIGN: The study is retrospective, and the data comprised 826 AGFs, intraoperatively measured in 111 patients implanted with a HiRes90K cochlear implant (Advanced Bionics). For each AGF, the eCAP threshold was determined using two commonly used methods: linear extrapolation (LE) toward the x axis and detection of the last visible (LV) eCAP. Subsequently, the threshold confidence interval (TCI) of each eCAP threshold was calculated to serve as a metric for precision, whereby a larger TCI means a lower precision or reliability. Additionally, the eCAP thresholds results were compared with most recent behavioral fitting thresholds (T profile) to put the eCAP threshold analysis in clinical context. Thereby, the association between eCAP and behavioral thresholds was calculated, both for all subjects together (group analysis) and, in contrast to previous studies, within individual subjects. RESULTS: Our data show that the TCIs were larger with the LE method than with the LV method. The eCAP thresholds estimated by the LE method were systematically smaller than those estimated by the LV method, while the LE thresholds with the smallest TCIs correlated best with the LV thresholds. Correlation analysis between eCAP and behavioral thresholds revealed correlation coefficients of r = 0.44 and r = 0.54 for the group analysis of LE and LV thresholds, respectively. Within individual subjects, however, the correlation coefficients varied from approximately -1 to +1 for both LE and LV thresholds. Further analysis showed that across subjects, the behavioral thresholds fell within the TCIs of the eCAP threshold profiles. CONCLUSION: This study shows that eCAP thresholds have an uncertainty that can be estimated using TCIs. The size of the TCI depends on several factors, for example, the threshold estimation method and measurement conditions, but it is often larger than one would expect when just looking at the threshold values. Given these large TCIs, future research on eCAP thresholds should be accompanied by a measure of precision to correctly apply eCAP thresholds in clinical practice. Comparing our eCAP threshold results with T profiles indicates that the eCAP thresholds are possibly not precise enough to predict T profiles.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Biesheuvel JD, Briaire JJ, Frijns JHM, et al. The Precision of eCAP Thresholds Derived From Amplitude Growth Functions. Ear Hear. 2018. PMID: 29219858. doi:10.1097/AUD.0000000000000527

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。